12.0
SERVICE
PARAMETERS
The scale will allow entry to the parameters if the
[Tare]
key is pressed during the power on
cycle. The passwords work as above.
In this case the display will show the passcode request screen,
“ P
- - - -
“
. To continue enter a
passcode as described below.
Entering passcode 0000 will allow calibration as shown in section 11.
Entering 1000 will allow access to a limited set of parameters described in section 12.1.
USING
“1000”
TO
ENTER
THE
SERVICE
PARAMETERS
Press the
[
➔
0/T
]
key during the display counting when turned on,
When
“Pn “
is displayed, enter the number 1000 using the
[Unit/
]
and
[Pcs/
]
keys and
then press
[
➔
0/T
]
.
The displays will show the first parameters, called
“F4 Int”
.
To select another parameter press the
[Pcs/
]
key to advance through the parameters.
Press the
[
➔
0/T
]
key to enter a parameter.
To exit a parameter, press the
[Print/M+/Esc]
key.
The display will show the parameter number and a name.
When a parameter is entered by pressing the
[
➔
0/T
]
key, the displays will guide you
through the parameter selected and the options available.
The parameters available are:
“F4 Int”
Initial Zero Range
“F5 rEZ”
Re-Zero range
“F6 SCS”
Successive Tare Enable
“F7 Cnt”
Display the A/D counts
“F8 Zem”
Zero Mode
“F9 Lvd”
Low voltage detection
“FA AZn”
Auto Zero Range
“Fb FPS”
User Function Password
The description of the parameters is shown in section 12.2.