background image

6 \ USER OPERATION MANUAL: 

Cole-Parmer Basic Ductless Fume Hood Series

 

 

 

 

 

4.

  Removing the Packaging Material. 

» The cabinet is protected by

 

 

extruded polystyrene foam, cardboard and/or shrink-wrap. 

» If you find any damage during this stage of unpacking please refer to the 

Damaged Freight Claim Information 

on our website. 

» We recommend leaving the cabinet secured with straps to the pallet until the cabinet is located in its 

approximate final position to facilitate ease and safety in handling. 

Note: Choosing the best location for your cabinet in order to achieve optimum operating performance 
is determined by a number of factors. Please refer to the next chapter for some guidelines. 

5.

  Moving the Cabinet. 

» When lifting the pallet with the cabinet, always ensure that the floor jack or mechanical lift truck has fully 

entered under the pallet. This is to increase the stability of the cabinet and reduce the risk of the cabinet 
falling down. Please use a suitable extension bar when necessary. During the moving of the cabinet, 
ensure there is enough distance between the supports of the pallet and the ground. Dragging the pallet 
against the ground will damage the pallet and possibly your new cabinet. 

» When removing cabinet from pallet or placing cabinet onto pallet, use at least two people. 

 

 

 

6.

  Removing the Strapping. 

» Remove the strapping by cutting it at a safe position to prevent any scratching the surface of your new 

cabinet. 

» Do not discard the packaging material for your cabinet until you have checked all of the components, installed 

and tested the unit. 

7.

  Lifting the Cabinet. 

» The cabinet can be lifted in two sections: The HEAD unit and ENCLOSURE. 

» Install the cabinet on the existing work surface or Cole-Parmer support stand (if ordered). 

Note: 

» When installing the cabinet onto an existing work surface, ensure that the structure can safely support the 

combined weight of the cabinet and any related equipment. Some modifications to the work surface may be 
necessary. 

» The work surface should be smooth, non-porous and resistant to the disinfectants and chemicals used 

in conjunction with the cabinet. 

 

2.2

  Packaging Contents

 

The following items are included with your manual: 

 

Test Report 

In case this manual and/or test report is lost or misplaced, the factory retains a copy in our files. A replacement 
copy can be obtained by contacting Cole-Parmer and providing the cabinet model, serial number and a brief 
description of the information desired. 

Содержание 78900-00

Страница 1: ...me Hoods USER OPERATION MANUAL l Revision No PURAIR BASIC SERIES V4 2020 pictured Model P5 XT with optional velomete Specifications are subject to change without notice or obligation 78900 00 78900 01...

Страница 2: ...Importance of Performance Validation Certification 16 3 7 Disclaimer 16 IV Operating Your Cabinet 17 4 1Control System 17 4 2 CabinetOperating Procedure 18 V Monitoring 19 5 1 General 19 5 2 Manual M...

Страница 3: ...hemical radiological or other non microbiological hazards are being used in the cabinet additional protective measures should be taken Additionally the operation should be monitored by a suitably trai...

Страница 4: ...egree of protection to the indicated service center To ensure prompt handling the return authorization number must be placed on the outside of the package A detailed explanation of the defect should b...

Страница 5: ...g further 2 1 Step By Step Procedure 1 Inspecting the Crate Pallet Boxes Upon receipt of your new cabinet inspect all cartons If there is any visible damage to the exterior please refer to Damaged Fre...

Страница 6: ...against the ground will damage the pallet and possibly your new cabinet When removing cabinet from pallet or placing cabinet onto pallet use at least two people 6 Removing the Strapping Remove the str...

Страница 7: ...e face velocity of the filter the potential exists for contaminated air to enter the work zone of the cabinet A minimum distance of 50 cm 20 in to the top of the ceiling is recommended for blower chan...

Страница 8: ...r Cabinet 1 Please refer to Unpacking Your Cabinet page 5 2 Inspect your cabinet carefully Should you find any defect please contact customer service 3 Peel off any protective masking that was left on...

Страница 9: ...3 Appropriate number of personnel are on hand two people recommended Fume Hood Assembly 1 Unpack fume hood enclosure and set on top of casework In most cases the enclosure is shipped fully assembled...

Страница 10: ...es of the cutout Two tabs are bent at 90 to allow it to hinge down into the workzone Once installed lay the pre filter on top of the tray 4 Main filter now requires a painted metal hat frame to be pla...

Страница 11: ...tack the second filter on top of the first with the gasket side down Make sure the corners are squared up to ensure proper installation 8 Place the white metal hat frame on top of the filter 9 Rotate...

Страница 12: ...nut and set aside DO NOT DISCARD 12 Remove any tape on the airflow sensor which is used to protect the sensor wires Install the sensor by attaching to the enclosure with the screw and nut 13 If not al...

Страница 13: ...f meter just below the screw b Carefully remove vane from plastic bag and cardboard envelope two vanes are enclosed one is a spare Hang the vane by the wire in the two slots provided in the vane holde...

Страница 14: ...replaced with the spare vane The vaneometer is accurate to 5 of full scale from 0 100 fpm and 10 from 100 fpm to 400 fpm The permanent mounting bracket provided in the box is not used 16 Removable yel...

Страница 15: ...e the blanking panel currently installed on the side of the fume hood 2 Remove the mounting screws in each corner of the blanking panel and disconnect the panel from the fume hood 3 Align the waste ch...

Страница 16: ...rmance Validation Certification An airflow velocity value that falls below the value specified inside the test report will not provide adequate operator protection 3 7 Disclaimer The performance of th...

Страница 17: ...tch up to turn on The green indicator lamp will illuminate to confirm the power is on The fan will start and the internal task light unit will illuminate Low Airflow Alarm The alarm status is displaye...

Страница 18: ...ring the filter block may temporarily reduce the efficiency of the filter For this reason any major spillage must be cleared up immediately preferably using spillage absorption granules rather than pa...

Страница 19: ...ct operating height using an anemometer such as a hot wire vane anemometer or propeller type Depending on the size of the cabinet a series of readings are to be taken at the front opening these are to...

Страница 20: ...d also be noted that fume cabinets are sometimes used to contain and protect users of the cabinet from hazardous or harmful substances Before commencing this schedule it is important to ensure the cab...

Страница 21: ...tial pressure switch to detect a high vacuum situation when the pre filter is blocked or blocking up The pressure switch is calibrated and tested prior to leaving our factory and under normal circumst...

Страница 22: ...ialist company NOTE CONSULT YOUR SAFETY OFFICER OR INDUSTRIAL HYGIENIST BEFORE REMOVING OR DISPOSING ANY FILTERS Pre Filters The pre filter is located below the main filter Remove the perforated pre f...

Страница 23: ...ly the front and top of the cabinet to remove any accumulated dust When needed use soap or other household mild detergent Quarterly 1 Replace pre filters 2 All monthly activities Annually 1 Replace al...

Страница 24: ...heck Low Airflow Alarm Check airflow at aperture Check pre filter is not blocked Check fan is running Recalibrate Filter Saturated Optional Check filter condition with Gastec or Draeger test kit Check...

Страница 25: ...RSONNEL WARNING Ensure main power supply has been removed prior to any work being carried out Light Units To change a light unit use the following instructions 1 Ensure the power cord has been disconn...

Страница 26: ...binations to best suit your specific application HEPA filters are available for applications involving particulates and can be combined together with any of our activated carbon filters 7 1 Filter Des...

Страница 27: ...27 USER OPERATION MANUAL Cole Parmer Basic Ductless Fume Hood Series VIII Product Specifications...

Страница 28: ...Ductless Fume Hood Series Cole Parmer 625 East Bunker Court Vernon Hills IL 60061 1844 Ph 1 800 323 4340 Fax 1 847 247 2929 www coleparmer com info coleparmer com 78900Basic_rev_1 0 2020 Cole Parmer...

Отзывы: