5.3. E-Cell commissioning
5.3.1. E-Cell module communication
To ensure that there is a working communication between the display unit and the electrochemical
cell module, follow the steps below using the display.
5.3.1.1. Press
to return to the main display.
From the main display press
to reach Diagnostics. Press
to access Diagnostic
data.
.
5.3.1.2. Use the
key to reach E-C CH1 Diag, and press
to access.
5.3.1.3. Use the
to reach DRx, and press
to access.
5.3.1.4. DRx: Data figures should appear on the screen as shown below. This indicates that the
display module is communicating with E-Cell module and is operating correctly.
5.3.1.5. With communication established, the following section lays out how detector levels can
be found.
5.3.2. Detector levels
In zero conditions the E-Cell detector levels should be around 26215 for CO (cell 1) and NO(cell 1),
32767 for NO2(cell 1). These detector levels can be found by using the steps below.
In order to access the detector level reading for E-
Cell 2 & 3, follow the same pathway, changing ‘E-C
CH1 Diag’ to ‘E-C CH2 Diag’ and ‘E-C CH3 Diag’.
1. Meas - 09:24:29
CO: 21 ppm
3. DIAGNOSTICS
3.2. E-C CH1 Diag
DRx
26200
3. DIAGNOSTICS
Содержание Tunnel Tech 700 Series
Страница 2: ...Issue A Rev 0 Date 25 04 17 Ref 100303...
Страница 3: ...CODEL Issue A Rev 0 Date 25 04 17 Ref 100303...
Страница 5: ...CODEL Issue A Rev 0 Date 25 04 17 Ref 100303...
Страница 6: ......
Страница 10: ...Technical Manual CODEL Issue A Rev 0 Date 25 04 17 Ref 100303...
Страница 21: ...Technical Manual CODEL 11 Figure 7 Electrical Schematic...
Страница 23: ...Technical Manual CODEL 13 Power Control Display PCB Electrochemical Cell PCB Communications...
Страница 24: ...Technical Manual CODEL 14 Visibility PCB Ma Output PCB S Top mA RLY 4 5 Bottom mA RLY 1 2...
Страница 37: ...Technical Manual CODEL 27 Alignment Laser Please Call CODEL mA Relay Card 803 604D...
Страница 79: ......
Страница 80: ......
Страница 81: ...Technical Manual Page 1 CODEL...