Technical Manual
CODEL
18
5.2.4. Calibration Vis
To begin calibration of VIS, follow the steps below on the display unit.
Note: When ‘Enter to CAL’ has been selected, allow 5-10 minutes for calibration to complete.
5.2.4.1. Press
to return to the main display.
From the main display press
to reach Setup Mode. Press
to access Setup
options.
.
5.2.4.2. If a Passcode is set, input the passcode and press
to access Setup Mode.
5.2.4.3. Use the
key to reach Calibration, and press
to access.
5.2.4.4. Use the
to reach VIS CH ZERO CAL, and press
to access.
5.2.4.5. VIS CH ZERO CAL: When this menu has been reached, press
and the system will
initiate a zero cal.
Note: Ensure that the signal path is clear.
1. Meas - 09:24:29
CO: 21 ppm
4. Setup Mode
4.1. CALIBRATION
VIS CH Zero CAL
Enter to CAL
4. Setup Mode
Enter PassCode
0x0000
Содержание Tunnel Tech 700 Series
Страница 2: ...Issue A Rev 0 Date 25 04 17 Ref 100303...
Страница 3: ...CODEL Issue A Rev 0 Date 25 04 17 Ref 100303...
Страница 5: ...CODEL Issue A Rev 0 Date 25 04 17 Ref 100303...
Страница 6: ......
Страница 10: ...Technical Manual CODEL Issue A Rev 0 Date 25 04 17 Ref 100303...
Страница 21: ...Technical Manual CODEL 11 Figure 7 Electrical Schematic...
Страница 23: ...Technical Manual CODEL 13 Power Control Display PCB Electrochemical Cell PCB Communications...
Страница 24: ...Technical Manual CODEL 14 Visibility PCB Ma Output PCB S Top mA RLY 4 5 Bottom mA RLY 1 2...
Страница 37: ...Technical Manual CODEL 27 Alignment Laser Please Call CODEL mA Relay Card 803 604D...
Страница 79: ......
Страница 80: ......
Страница 81: ...Technical Manual Page 1 CODEL...