CODAN Infusion device training module & competency assessment
Quality · Reliability · Precision
User Initials…………..
Page 15 of 29
3. Standard Operation
3.5 Checking the Infusion Information
♦
User Aim – Demonstrate the ability to check the Infusion information
3.6 Changing the Rate whilst the Pump is Stopped
♦
User Aim – Demonstrate the ability to change the rate whilst
the pump is stopped
Stop the running infusion
Press the Clear keys
which will reset the rate to zero
Enter a new rate
Restart the infusion at the new rate
3.7 Utilise Key Lock
♦
User Aim – Demonstrate the ability to use the key lock
Whilst the pump is running press the menu key
until the key lock menu is reached
Enter the Trusts standardised code using the corresponding numeric keys.
Once the code has been entered confirm using the start/stop/enter key. A padlock will appear in
the bottom left of the display screen.
Return to your main screen using the exit key.
Repeat process to disable key lock
Press the Info Key
Row 1: Amount & time of total infusion
Row 2: Infused volume and elapsed infusion time
Row 3: Remaining volume and estimated time until the
end of infusion