CODAN Infusion device training module & competency assessment
Quality · Reliability · Precision
User Initials…………..
Page 13 of 29
3. Standard Operation
♦
User Aim – For the user to load the pump and set up an infusion
3.1 Infusion Set Selection
Press ON/Off: the pump is switched on.
The infusion set menu is automatically called up.
Press the selection key to enter the desired infusion set type.
Press the Enter key to confirm the parameters.
3.2 Priming and Loading the Giving Set
Ensure you have the correct CODAN giving set.
Open the package and ensure the roller clamp is closed.
Spike the bag/bottle/container and fill the chamber to approx half way.
Open the roller clamp and allow the fluid to progress to the end of the line.
REMEMBER
ALL CODAN SETS CONTAIN A FLOW STOP CAP
Manually close the “Anti free flow” clamp.
Insert the set right across the middle of the pump ensuring the “Anti free flow clamp” is inserted
correctly into its housing as shown:
Close the door.
Open the roller clamp.
You are now ready to use the CODAN Argus 718V.