CODAN Argus 718V Infusion Pump

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

A Step By Step Training Module 

 

NAME: ..............................................

 

 
 

SOUTHERN HEALTH

 

Workbook 718V, Rev 01, Nov 2018 User Manual: 14.547_G_UM_en_A71XV_5.09 

End User Workbook & 

 Competency Assessment 

Содержание Argus 718V

Страница 1: ...CODAN Argus 718V Infusion Pump A Step By Step Training Module NAME SOUTHERN HEALTH Workbook 718V Rev 01 Nov 2018 User Manual 14 547_G_UM_en_A71XV_5 09 End User Workbook Competency Assessment...

Страница 2: ...you have gained the required standard please inform your lead nurse who will be happy to take you through your Competency Assessment Good Luck Contra indication The ARGUS 718 V is not designed and ap...

Страница 3: ...n 3 2 Priming and Loading the Giving Set 3 3 Infusion Specifying the Flow Rate and Volume 3 4 Infusion Specifying the Volume and Time 3 5 Checking the Infusion Information 3 6 Changing the Rate whilst...

Страница 4: ...er Initials Page 4 of 29 1 External features of the CODAN Argus 718V User Aim To become familiar with CODAN Argus 718V external features Front 1 Indicator for mains operation Battery recharge 2 Status...

Страница 5: ...eatures of the CODAN Argus 718V User Aim To become familiar with CODAN Argus 718V external features Rear 1 Power connection 2 DC connection interface for a medical PC 3 Nurse call 4 Optional interface...

Страница 6: ...ing module competency assessment Quality Reliability Precision Page 6 of 29 External features of the CODAN Argus 718V User Aim To become familiar with CODAN Argus 718V external features Front showing...

Страница 7: ...g module competency assessment Quality Reliability Precision User Initials Page 7 of 29 External features of the CODAN Argus 718V User Aim To become familiar with CODAN Argus 718V external features Si...

Страница 8: ...ce training module competency assessment Quality Reliability Precision User Initials Page 8 of 29 External features of the CODAN Argus 718V User Aim To become familiar with CODAN Argus 718V external f...

Страница 9: ...ime The keys correspond to the numbers of the display starting from the right The value of the corresponding number is increased by one unit by repeatedly pressing one of these keys Clear Function Pre...

Страница 10: ...t your needs Certain menus are active only immediately after the pump is switched on or an infusion set has been changed For safety reasons these menus are deactivated after an infusion is first start...

Страница 11: ...t value is exceeded the unit triggers a pressure alarm Each time the pump is switched on again the occlusion pressure limit is automatically reset to the preconfigured default value Key Lock For all m...

Страница 12: ...t is connected to the mains Unit is on battery power Unit connected to a CODAN Argus Docking Station Key Lock Total volume or total infusion time Previously infused volume or previous infusion time Re...

Страница 13: ...the parameters 3 2 Priming and Loading the Giving Set Ensure you have the correct CODAN giving set Open the package and ensure the roller clamp is closed Spike the bag bottle container and fill the c...

Страница 14: ...used of 500ml Press start to begin your infusion 3 4 Infusion Specifying the Volume and Time User Aim Demonstrate the ability to set up an infusion using Volume to be infused and Time Press the select...

Страница 15: ...a new rate Restart the infusion at the new rate 3 7 Utilise Key Lock User Aim Demonstrate the ability to use the key lock Whilst the pump is running press the menu key until the key lock menu is reach...

Страница 16: ...audible signal can be configured After the pre alarm is acknowledged the status light changes to amber and remains on until the cause of the pre alarm is resolved or the pre alarm is replaced by an al...

Страница 17: ...y means of an icon and text In any event the error must be eliminated before restarting the infusion For further information see Section 15 2 Alarms in User Manual 14 547_G_UM_en_A71XV_5 09 Procedure...

Страница 18: ...The battery has completely depleted Connect to the mains Pump Unused The pump has been left on or in stop mode for a set period of time The time can be configured between 1 60 mins Air Alarm The size...

Страница 19: ...e contact your EBME Technical service dept Stop Pump First The configuration determines the pump must be stopped first Locked Entry Not Allowed The Key lock has been activated Infusion without Drop De...

Страница 20: ...ee Flow Clamp The wrong set or a set with an AFF has been inserted Insert Infusion Set First There is no giving set in the pump Remove Air Bubbles First There are still air bubbles in the giving set C...

Страница 21: ...d Competency Outcome Competency Achieved Mentor Assessor Date Can demonstrate equipment required to set up the pump CODAN Argus 718V infusion pump CODAN giving sets Red roller clamp for blood and Blue...

Страница 22: ...arity with the external components of the volumetric pump Display panel Alarm indicator Door handle Upstream pressure sensor Downstream pressure sensor Air bubble detector Housing for anti free flow c...

Страница 23: ...ctively the functions of the pump display panel Operational symbols Display panel mls hr Alarm indicators Menu key Alarm silence key Start Stop key Numerical keys On Off key Can demonstrate effective...

Страница 24: ...Revisited Competency Outcome Competency Achieved Mentor Assessor Date Can safely discontinue the pump Monitor the patient s condition observations and ensure their condition is stable prior to discont...

Страница 25: ...t must be immediately disconnected from the mains or removed from the docking station The pump must be dried immediately and thoroughly cleaned If there is any likelihood of fluid ingress the unit mus...

Страница 26: ...rvice Department 2 Infusion Lines Only standard infusions sets with LUER lock connections A708V A718V and Anti Free Flow Clamp which have been validated and approved by CODAN ARGUS AG are to be used T...

Страница 27: ...gus 718V is switched off but also after a complete power outage the setting parameters are preserved for a limited time 4 Event History Significant events are recorded in the pump s memory in chronolo...

Страница 28: ...rug Library Options Drug Library Dose Calculation Med DB Drug Error Reduction software The CODAN Argus 718V contains a simple drug library which holds a maximum of 256 drugs Each library can be filter...

Страница 29: ...roducts Regulatory Agency www mhra gov uk The National Institute for Clinical Excellence NICE www nice org uk National Patient Safety Agency www npsa nhs uk Nursing and Midwifery Council www nmc uk or...