3. Operation
The
EZ
Test HazMon System Tester has an easy to use menu system. To turn on the tester, press
and hold the “
POWER
” key until the display backlight illuminates. To turn the tester off, press and hold
the
“
POWER
” key until the display backlight goes off. The tester automatically turns off after 5 minutes
except under the following conditions:
1.
The tester is displaying real-time information from a sensor;
2.
The tester is in “Bus Test” mode (the tester must be powered in “Bus Test” mode or it will
cause the network to fault when connected).
The system will display two opening messages when power is turned on:
The tester may display a status message before proceeding to the main menu. The status messages
are as follows:
The main menu is then displayed:
Message
Description
Required Action
Battery Low
< 25%
The battery has
discharged below 25%
Check the battery level using the
“Battery Gauge” function. If less than
10% replace the batteries
Program CRC
Fault
The tester’s program is
corrupted
Return the unit to the factory for repair
Data RAM
Fault
The tester failed self-test
Return the unit to the factory for repair
Calibration
Fault
The calibration
parameters are corrupted
Return the unit to the factory for repair
CMC Industrial
Electronics
EZTest Ver 1.00
Copyright 2012
Select Function
Sensor Test