background image

13/11/2018 

 

Page 11 

Operation 

1. Place  the  unit  on  a  sturdy  and  level  surface  in  a  safe,  dry  place, 

away from laboratory traffic.  

2. Ensure  that  the  AC  power  switch is  OFF,  and  then  plug  the  three-

pronged  power  cord  one  end  into  a  grounded  three-prong  AC 
outlet with appropriate voltage (100V to 240V as indicated on the 
rating sticker near the AC cord on the back of the unit) and plug the 
other end into the main power socket. 

3. Connect the DC output jacks from the electrophoresis unit; insert the 

red lead (+) into the red output jack, and the black lead (-) into the 
black output jack.  

4. Power on the unit by pressing the ON/OFF switch on the back. 

5. The screen will show a loading screen for few seconds then enter to 

the mode-selecting page. 

Constant Mode 

Use the 

Constant  Voltage  /  Current  / Power  Operation 

for applications that 

require  only  one  specific  voltage  limit,  current  limit,  and  power  limit 
continuously during the entire operation of electrophoresis. 

Note:  When  target  constant  mode  is  set,  system  automatically  adjusts  the  other  two 
parameters  to  maximum  to  allow  constant  run.  For  example,  if  constant  voltage  is  set, 
system will adjust current and power to the maximum value.  

Users could later lower the other two values by themselves. System will hold the value either 
at target constant value or the one which has been reached first 

1. Use the up and down arrow keys to select 

constant mode, then press the enter key to 
move to the next page. 

Содержание powerPRO Series

Страница 1: ...POWERPRO...

Страница 2: ......

Страница 3: ...PRO500 powerPRO3AMP Record the following for your records Model _____________________ Catalogue No _____________________ Date of Delivery _____________________ Warranty Period _____________________ Se...

Страница 4: ...uipment Operation 5 Symbols 5 Potential Risk and Preventive Measures 6 Risk assessment table 6 Preventative measures 6 Packing List 7 Specifications 8 Operating Instructions 10 Installation 10 Control...

Страница 5: ...ore operating this equipment 1 Read and follow the manual instructions carefully 2 Do not alter the equipment Failure to follow these directions could result in personal and or laboratory hazards as w...

Страница 6: ...ation of the unit The powerPRO Power Supply has been designed to utilize shielded wires thus minimizing any potential shock hazard to the user Cleaver Scientific recommends against the use of unshield...

Страница 7: ...uction to see if the proposed method will damage the equipment Equipment Operation Follow the guidelines below to ensure safe operation of the unit 1 NEVER access dangerous chemicals or other material...

Страница 8: ...500 V Max current 800mA Max watt 300W Potential Risk and Preventive Measures Risk assessment table Frequency Risk Frequent Likely Possible Rare Unlikely Bruise Slash Electrical shock Power cord plug w...

Страница 9: ...acking lists should be referred to as soon as the units are received to ensure that all components have been included The unit should be checked for damage when received Cleaver Scientific is liable f...

Страница 10: ...be changed by user Program Storage 30 programmed files Program Multi Step Up to 6 steps Editable Program Function 1 Have typical running conditions program 2 Manual editable program Display 2 4 TFT Co...

Страница 11: ...Furthermore the powerful specifications plus five terminal pairs can be used for multi electrophoresis units simultaneously Cleaver Scientific powerPRO Series Power Supply provides Constant Voltage or...

Страница 12: ...the unit Press to move cursor left between parameters Pause Key Press to temporarily interrupt power to an operation in progress resume power after pausing without resetting the timer Press to move c...

Страница 13: ...witch on the back 5 The screen will show a loading screen for few seconds then enter to the mode selecting page Constant Mode Use the Constant Voltage Current Power Operation for applications that req...

Страница 14: ...lier will constantly operate until user manually stops it 5 Press the start Key to begin electrophoresis The LED next to the start key will light indicating the power supply is active Parameter values...

Страница 15: ...he operation set the Program Mode as follow 1 Use the up and down arrow keys to select program mode then press the enter key to move to the next page 2 Use the up and down arrows to select the desired...

Страница 16: ...run 8 Upon completion of the run an alarm will sound and the screen will display Complete 9 Press the start key to terminate the run and press the setting key to return to the settings screen Typical...

Страница 17: ...gories that can be selected Choose the model according to your experiment need 3 After selecting model the pre set value will display These values can be adjusted as described previously 4 Press the s...

Страница 18: ...chine back to local distributor or our Service Department for maintenance Operation stops with alarm The screen displays Electrophoresis leads are not connected to the power supply or to the electroph...

Страница 19: ...arm Display shows Ground leak detected during run Check the electrophoresis system for improper grounding Restart the power supply by turning the Power switch off and on Operation stops with alarm Dis...

Страница 20: ...nspect the fuses If the fuses are burned or there is a break in the fuses element replace the fuses with identical type of fuses 2 Place the fuses holder back into the compartment 3 Snap the cover clo...

Страница 21: ...anyone other than CSL or an appointed distributor or representative are no longer under warranty from the time the unit was modified Units which have accessories or repaired parts not supplied by CSL...

Страница 22: ......

Страница 23: ......

Страница 24: ......

Отзывы: