13/11/2018
Page 17
Operation stops with
alarm:
Display shows
Circuit is interrupted
Verify that the running buffer is correct.
Verify the all cables are attached
correctly
Turn the Power switch off and on again;
restart application.
If you cannot restart the instrument, turn
off the power, disconnect the power cord
from the outlet, and contact Technical
Service.
Operation stops with
alarm:
Display shows
Circuit is interrupted
Verify that the running buffer is correct.
Verify the all cables are attached
correctly
Turn the Power switch off and on again;
restart application.
If you cannot restart the instrument, turn
off the power, disconnect the power cord
from the outlet, and contact Technical
Service.
Operation stops with
alarm:
Display shows
Ground leak detected
during run
Check the electrophoresis system for
improper grounding. Restart the power
supply by turning the Power switch off and
on.
Operation stops with
alarm:
Display shows
Power supply is
overheating
Turn off power supply. Check for sufficient
airflow around the power supply fan. After
cooling down, restart the power supply by
turning the Power switch to the on
position.
If you cannot restart the instrument, turn
off the power, disconnect the power cord
from the outlet, and contact Technical
Service.
Warning message
displays with 5-second
beep sound.
The screen shows
The power once been cut
and now recover
User does not need to take extra action.
The warning sign and beep sound would
only last for 5 second; after that, the
machine will continue running the
unfinished project.
The
sign indicates the machine has
been interrupted by sudden power off.
Press Enter Key
to clear the sign.