Cleaver Scientific CVS10CBS Скачать руководство пользователя страница 23

 

 

23 

 

Problem: Electrophoresis

 

Cause

 

Solution

 

Current zero or less than 
expected and samples do not 
migrate into gel

 

Tape at the bottom of precast 
gel cassette not removed 
Insufficient buffer in inner buffer 
chamber 
Insufficient buffer in outer buffer 
chamber 
Electrical disconnection

 

Remove tape. 
 
Fill buffer chamber with running buffer. 
 
Fill inner and outer buffer chambers to ensure 
wells are completely covered. 
Check electrodes and connections.

 

Gels run faster than expected 

Running buffer too concentrated 
and gel temperature too high; 
incorrect running buffer 
concentration or type used 
Running or reservoir buffer too 
dilute 
Voltage too high 

Check buffer composition and type. 
 
 
 
Check buffer protocol and concentrate if 
necessary. 
Decrease voltage by 25-50%. 

Gels run slower than expected 

Incorrect running buffer 
composition or type 
Excessive salt in sample 

Check buffer composition and type. 
 
Desalt sample. 

Problem: Total Protein 
Staining

 

Cause

 

Solution

 

Bands not visible

 

No protein in gel 
 
Imaging system malfunctioning 
 
Incorrect imaging parameters 
were used

 

Stain with another method to confirm there is 
protein. 
Check instrument manual for troubleshooting 
or contact imaging instrument manufacturer. 
Check Instrument manual.

 

Poor staining sensitivity 

Dirty staining trays 
 
Insufficient stain volume 
 
Insufficient staining time 
Reuse of staining solution 

Clean staining trays and other equipment 
with laboratory glassware cleaner. 
Follow recommendations for stain volume 
(appropriate to gel size). 
Increase staining time. 
Repeat staining protocol with fresh staining 
solution. 

High or uneven background 
staining 

Staining trays or equipment dirty 
 
Too much time in staining 
solution 
 
 

 
Reagent impurities 

Clean staining trays and other equipment 
with laboratory glassware cleaner. 
Restrict duration of incubation in staining 
solutions as recommended in protocol. 
Wash gel in water or retrospective destaining 
solution for >30min. 

Use high-purity water and reagents for 
staining. 

Speckles or blotches in gel 
image 

Particulate material from 
reagents, staining tray, dust or 
gloves 

Clean staining trays thoroughly. 
Decrease time that gels and staining solution 
are exposed to open air. 
Use dust-free gloves and handle gels only by 
edges. 

Uneven staining 

Insufficient shaking during 
staining 

Agitate gel during staining. 

Gel shrinkage 

Gel dehydrated 

Transfer gel to water for rehydration. 

Problem: Evaluation of 
Separation

 

Cause

 

Solution

 

Diffuse or broad bands

 

Poor quality acrylamide or bis-
acrylamide incomplete 
polymerization 
Old SDS or sample buffer 

Use electrophoresis-grade reagents. 
Check polymerization conditions. 
 
Prepare fresh solutions. 

Содержание CVS10CBS

Страница 1: ...gue Numbers CVS10CBS SB10 Record the following for your records Model _____________________ Catalogue No _____________________ Date of Delivery _____________________ Warranty Period __________________...

Страница 2: ...e Guidance and restrictions 6 Setting up the Gel Tanks 6 Protein Electrophoresis 7 Casting Unit Preparation 7 Gel Preparation 9 Preparation of denatured protein samples for loading 11 Gel Pouring 11 G...

Страница 3: ...tly in position The unit should not be used if there is any sign of damage to the external tank or lid Acrylamide is a powerful neurotoxin in solution form Polymerized gels can contain some unpolymeri...

Страница 4: ...g List Checked by ________________________ Date ________________________ The packing lists should be referred to as soon as the units are received to ensure that all components have been included The...

Страница 5: ...de dishwashing liquid Hexane and Aliphatic hydrocarbons The units should not be left to in detergents for more than 30 minutes The units should never come into contact with the following cleaning agen...

Страница 6: ...ionally however a temporary conductivity caused by condensation must be expected Setting up the Gel Tanks Note Before setting up the Gel Tank please ensure that it has been properly cleaned and dried...

Страница 7: ...ause leaks Gel Cassette Assembly Assemble the glass plates in the CVS10DRIM casting running insert so that the bottom of the glass plates and the spacers are perfectly aligned on a flat levelled surfa...

Страница 8: ...When using the Slide Clamp Mini version simply slide both gates outwards until fully tightened When only one gel is being run the dummy plate must be used in the second position and fully tightened NO...

Страница 9: ...e clamps or screws as preferred by the user For those that prefer to use the screws rather than clamps the screws can be simply inserted into the screw holes The clamps can be removed by placing each...

Страница 10: ...olumes of solutions from the standard stock solutions These should be gently mixed avoiding generation of bubbles which will inhibit polymerization by removing free radicals Resolving Gel for 2 x 1mm...

Страница 11: ...combine the protein sample and 4 X sample buffer It is always advisable to use protein markers in one of the end lanes to indicate sizes of bands These should be prepared according to the manufacture...

Страница 12: ...n this use fresher stock solutions or add more APS and TEMED 7 Prepare the stacking gel solution 8 Before casting the stacking gel insert a piece of filter paper to dry the area in between the glass p...

Страница 13: ...ideways with the end s of the tank and pressed into the recess Or these can be fitted down the front of the tank Note NEVER FIT THESE UNDERNEATH THE MODULE IN THE BOTTOM OF THE TANK AS THIS WILL PREVE...

Страница 14: ...ve gel level Cooling packs are inserted behind the gels Outer Tank is filled to the maximum fill line Cooling is at a maximum 1000ml Gel Running 1 Fit the lid and connect to a power supply 2 Gels shou...

Страница 15: ...door Twist key applying pressure to both the clamping door and the CVS10D side cheek The door will now click open Repeat this process until you have opened both the doors 4 Remove the glass plates The...

Страница 16: ...and trapped bubble should be removed with a roller at each step during blotting sandwich assembly 1 Each blot sandwich should be set up according to the following order Cassette clamp ve black side pl...

Страница 17: ...entation and order the blot sandwiches were loaded in This can be done by noting which samples were loaded adjacent to each electrode and by marking the membrane with a pencil on one corner 2 Use of a...

Страница 18: ...he membrane is now ready to be probed Suggested Running Conditions please note that these are just guideline Duration of Blot SB10 One Hours Three Hours Overnight 16 hr 100V 125V 250mA 400mA 50V 200mA...

Страница 19: ...o 200ml with Distilled Water Stock 4 X Stacking Tris 0 5 M Tris HCL pH6 8 0 4 SDS To 110ml Distilled Water add 12 12 g of Tris base Add 8ml of 10 SDS Adjust pH to 6 8 with 1N HCl Add Distilled Water t...

Страница 20: ...03 g Bromophenol blue Aliquot into 1 5ml microcentrifuge tubes Store at 20 C Towbin Buffer 25mM Tris 192mM glycine 20 methanol pH8 3 Towbin Buffer SDS 25mM Tris 192mM glycine 20 methanol pH8 3 0 05 0...

Страница 21: ...pH will vary according to the freshness of the reagents used References 1 Sambrook Fritsch and Maniatis Molecular Cloning A Laboratory Manual Second Edition Cold Spring Harbor Laboratory Press 1989 2...

Страница 22: ...0 06 APS and 0 12 TEMED Gel does not polymerize Too little or too much APS or TEMED Failure to degas Temperature too low Poor quality acrylamide or bis Old APS Use 0 0 05 APS and 0 05 TEMED Degas mono...

Страница 23: ...rument manual Poor staining sensitivity Dirty staining trays Insufficient stain volume Insufficient staining time Reuse of staining solution Clean staining trays and other equipment with laboratory gl...

Страница 24: ...ing gel Uneven gel interface Remove salts from sample by dialysis or desalting column prior to sample preparation Use same buffer in samples as in gel Check buffer composition and dilution instruction...

Страница 25: ...extent that it prevents binding to nitrocellulose Alternatively reduce methanol concentration and switch to PVDF Protein precipitating in gel Use SDS in transfer buffer SDS can increase transfer effi...

Страница 26: ...brane PVDF Membrane is not completely wet Because of hydrophobicity of PVDF the membrane must be soaked entirely in methanol before equilibration in aqueous buffer Proteins might be transferring throu...

Страница 27: ...27 Too low background Increase antibody concentration protein sample concentration Consult manual included with antibody detection kit Total protein detection Consult stain or detection kit manual...

Страница 28: ...ormed by anyone other than CSL or an appointed distributor or representative are no longer under warranty from the time the unit was modified Units which have accessories or repaired parts not supplie...

Отзывы: