Resistance
600.0
Ω
± (1%+5)
6.000k
Ω
± (1%+5)
60.00k
Ω
± (1%+5)
600.0k
Ω
± (1%+5)
6.000M
Ω
± (1.5%+5)
60.00M
Ω
± (2.5%+10)
Capacitance
6.000nF
± (5%+15)
60.00nF
± (3%+8)
600.0nF
± (3%+8)
6000nF
± (3%+8)
60.00µF
± (3%+8)
600.0µF
± (5%+8)
6.000mF
± (8%+8)
Diode
2.000V
Test Current: 0.5mA
Continuity
Buzzer beeps if the
resistance is less then
25
Ω
.
APPENDIX B
• The scopemeter does not be start up.
The battery may be completely empty. ln this case the scopemeter will not start up, even if it is powered by
the battery charger. Charge the battery first: power the scopemeter with the battery charger without turning it
on. Wait about 15 minutes and then try turning on the scopemeter again. If the problem remains, contact our
company.
• The scopemeter turns off automatically after operating for only several seconds.
Maybe the battery is empty. Please check the battery charge indicator on the right bottom of the screen· Charge
the battery if it is empty.
• How to Change between AC Voltage and DC Voltage measurements.
Press F2 when voltage measurement interface is disp!ayed.
• ln the scope mode, the measurement result is 10 times bigger or smaller than the actual value.
Check whether the channel A or channel B's probe attenuation coefficient setting in the scope interface is the
same as the switch setting of this probe; for examp!e, both the two setting are "x10" or "x1 ".