26
English
10.4 AC Current
Range
Resolution
Accuracy
Overload protection
45–1000 Hz
> 1–3 kHz
600 µA
0.1 µA
±(1.2 % + 5)
±(1.5 % + 5)
Fuse 1: F1A H 240 V (CE),
ø 6 × 25 mm
6000 µA
1 µA
60 mA
0.01 mA
±(1.5 % + 5)
±(2.0 % + 5)
600 mA
0.1 mA
6 A
1 mA
±(2.0 % + 5)
±(3.0 % + 5)
Fuse 2: F10A H 240 V (CE),
ø 6 × 25 mm
10 A
0.01 A
10.5 Resistance
Range
Resolution
Accuracy
Overload protection
Remark
600 Ω
0.1 Ω
±(1.2 % + 3)
1000 V DC / 750 V AC When
measuring
below 2 kΩ,
apply REL
to ensure
measurement
accuracy.
6 kΩ
0.001 kΩ
±(1.0 % + 2)
60 kΩ
0.01 kΩ
600 kΩ
0.1 kΩ
6 MΩ
0.001 MΩ
±(1,5 % + 2)
60 MΩ
0.01 MΩ
10.6 Capacitance
Range
Resolution
Accuracy
Overload protection
Remark
40 nF
0.01 nF
±(3.0 % + 5)
1000 V DC / 750 V AC There is around
10 nF residual
reading when
the circuit
is open.
400 nF
0.1 nF
4 µF
0.001 µF
40 µF
0.01 µF
400 µF
0.1 µF
±(4.0 % + 5)
4000 µF
1 µF
unspecified
Содержание ut61d