10
English
Note:
In each range, the meter has an input impedance of 10 MΩ except the mV range
which has an input impedance of 3000 MΩ. This can cause measurement errors
in high impedance circuits. If the circuit impedance is less than or equal to 10 kΩ,
the error is negligible (0.1 % or less).
5.1.2 Measuring AC voltage
1. Connect the black test lead
to
and the red test
lead to
.
2. Set the function selector
to
for AC voltage
measurement.
3. The meter is set by default
to
Autorange
(automatic
range setup). If you wish to
change the range manually,
press [ Range ] several times.
4. Connect the test leads
across the source or load to
be measured.
5. The true RMS value will be
displayed.
6. Press [ Hz % ] to obtain
the frequency and Duty
Cycle value.
Input amplitude:
≥ range × 30 %
Frequency response: ≤ 1 kHz
Note:
In each range, the meter has an input impedance of 10 MΩ except the mV range
which has an input impedance of 3000 MΩ. This can cause measurement errors
in high impedance circuits. If the circuit impedance is less than or equal to 10 kΩ,
the error is negligible (0.1 % or less).
Содержание ut61d
Страница 2: ...2...