Installation
10
CONNECTION
• Locate power cable connector at the rear of the generator.
• Before connecting the cable, ensure that the POWER key is set to OFF.
• Connect the power cable (length 2 m, type IEC350-Schuko) supplied with the
generator.
4.6.
Warnings
e
If a failure in gas supply (due to a power failure, activation of an electrical safety
device, or generator fault), even temporary, is not admissible, a pneumatic panel
should be envisaged, to enable provisional activation of a reserve gas source (to-
tally or partially automatic)
e
To ensure compliance with standards imposed by the Machinery directive, the
start-up of the generator after electrical power supply is restored cannot normal-
ly be automatic, and should be manually activated by the operator
e
If automatic start-up of the generator is necessary, this command should be en-
tered in the set-up program
4.7.
Packaging disposal
All packaging materials used are compatible with the environment and may be
stored without risk, or disposed of and recycled in accordance with local legisla-
tion governing classified waste collection.
However, when possible, it is recommended to store the packaging to ensure
adequate protection when moved in the future.