Maintenance
20
7.5.
Vent silencer filters
The vent silencers are fitted on solenoid valves which depressurise the molecular
sieves during regeration. The vent silencers are used to dampen the noise ge-
nerated by the high speed delivery of gas from the solenoid valve.
7.6.
Main fuse
The generator is protected by means of a fuse located in the electric power soc-
ket.
For replacement proceed as follows:
REPLACEMENT
• Turn off the generator and disconnect
the power cable; remove the front
lower panel by loosening the fixing
screws.
• Locate the position of the solenoid val-
ves, at the lower left and right of the
generator.
• Locate the cylindrical silencers
A
and
B
, placed vertically above the solenoid
valves.
• Turn the silencers anti-clockwise to lo-
osen and replace with new ones.
• Close the instrument and reconnect
the power cable.
• The generator can now be restarted.
• Remove the power cable from the soc-
ket
• Extract the fuse box
A
from the socket
• Replace the faulty fuse
• Refit the box and cable in the socket
• The generator can now be restarted.