Nitrogen Generator
NiGen GC Nigen MICRO
31
8.3.
List of warnings
8.4.
List of errors
warning
identification
cause
resolution
W-00
: data loss
the memory has lost the set
up values of the parameters
and has restored the default
values
contact service in order to
set the correct values of the
parameters
W-01
: low outlet pressure
outlet pressure below thre-
shold
reduce outlet flow
W-02
: maintenance requi-
red
maintenance counter has
decreased to zero
check maintanance sche-
dule; perform the reque-
sted maintenance operation;
restore related mainte-
nance counter according to
table at
par. 7.2.1.
W-03
: catalyzer warm-up
time over
the catalyzer has not rea-
ched the set-point tempera-
ture
contact service
W-04
: low catalyzer tem-
perature
the catalyzer can not main-
tain the set-point tempera-
ture
contact service
error
identification
cause
resolution
E-00
type error
generator not identified
assign a number to the
generator (
see Stack user’s
manual
)
E-01
low inlet pressure
inlet air pressure below
minimum operative value
increase inlet air pressure
E-02
high inlet pressure
nlet air pressure above
maximum acceptable value
decrease inlet air pressure
E-03
hight catalyzer tem-
perature
catalyzer temperature above
maximum acceptable value
contact service
E-04
communication fault
control board can not read
data
contact service