4.4.
Nitrogen outlet
m
The distribution line must be made of pipe rated for a pressure of at
least 10.5 bar (152 psi).
The pipe diameter must be sized on the basis of the acceptable pressure drop
values of the utility.
e
To ensure nominal purity of the delivered gas is maintained (minimize contami-
nation from oil residue, dust, scaling and humidity), it is recommended to use
clean and/or treated metal pipelines for final gas distribution.
CONNECTION
• Locate the nitrogen coupling, marked “NITROGEN OUTLET” at the rear of the
generator;
• take fitting (3) and mount it;
• connect the line.