 Plug the red test lead in  TEMP/mA  jack and the 

black test lead in 

 jack.

 Remove power from the circuit that is to be 

measured. Open up the circuit and 

connect the black lead to the nega-

tive side and the red lead to the 

positive side of the circuit so that 

the test leads are in series with the 

load to be measured.

 Apply power to the circuit. 

 Read the displayed current.

 If the numeral value in the display 

is too small, change the function 

switch to next lower range. Repeat 

until you have maximum digits pos-

sible without displaying ʻIʼ.

 If ʻIʼ appears on the display, it indicates over-range. Immediately remove 

test leads from the measuring circuit to avoid any damage to the meter. 

b) 200mA or more DC Current Measurement
CAUTION!

 Do not exceed 15 seconds when measuring the 20A range and 

wait for 30 minutes between each measurement. 

 Set the function switch to 20 on the 

DC Amps

 

scale.

 The LED above  20A  jack will turn on. 

 Plug the red test lead in  20A  jack and 

the black test lead in 

 jack.

 Remove power from the circuit that 

is to be measured. Open up the 

circuit and connect the black lead to 

the negative side and the red lead 

to the positive side of the circuit so 

that the test leads are in series with 

the load to be measured.

 Apply power to the circuit.

 Read the displayed current.

CAP

CIRCUIT-TEST

DMR-4200

20A

TEMP

mA

HOLD

POWER

CAT II 1000V

V/Ω/F/   /

COM

D

C

 V

O

LT

S

 

OHMS 

Ω

C

A

PA

C

IT

A

N

CE

AC

 V

O

LT

S

AC AMPS

D

C A

M

PS

TEMP F

REQ

+ DC –

CAP

CIRCUIT-TEST

DMR-4200

20A

TEMP

mA

HOLD

POWER

CAT II 1000V

V/Ω/F/   /

COM

D

C

 V

O

LT

S

 

OHMS 

Ω

C

A

PA

C

IT

A

N

CE

AC

 V

O

LT

S

AC AMPS

D

C A

M

PS

TEMP F

REQ

+ DC –

Содержание DMR-4200

Страница 1: ...USER S MANUAL DIGITAL MULTIMETER DMR 4200 CIRCUIT TEST ELECTRONICS www circuittest com...

Страница 2: ......

Страница 3: ...ent 8 3 DC Current Measurement 8 10 a less than 200mA DC current b 200mA or more DC current 4 AC Current Measurement 10 11 a less than 200mA DC current b 200mA or more AC current 5 Resistance Measurem...

Страница 4: ...C RMS as these volt ages pose a shock hazard Always consider circuits to be energized Never assume any equipment to be de energized Always start with power off Set the function switch to the correct s...

Страница 5: ...with respect to ground exceeds 500V AC DC This symbol adjacent to one or more terminals indicates them as being associated with ranges that may in normal use be subjected to particularly hazardous vo...

Страница 6: ...st Lead Input Indicating LEDs Lighted LED indicates the correct jack to plug in the test lead for the selected function LED will turn off after the test lead is fully inserted If the test lead is inco...

Страница 7: ...rent of 1 0 mA maximum Continuity Audible signal sounds if resistance is less than 30 Over range indication I is displayed Operating Temp 41 to 95 F 5 to 35 C Storage Temp 14 to 122 F 10 to 50 C Relat...

Страница 8: ...ading 1 digit 2 0 reading 5 digits AC CURRENT AC A 40 400HZ 2mA 200mA 20A 1 A 100 A 10mA 1 0 reading 3 digits 1 8 reading 3 digits 3 0 reading 7 digits RESISTANCE 200 2k 20k 200k 2M 20M 0 1 1 10 100 1...

Страница 9: ...REOFTHEVOLTAGE BEING MEASURED SELECT THE HIGHEST RANGE AND REDUCE UNTIL A SATISFACTORY READING IS OBTAINED Set the function switch to 1000 on the DC VOLTS scale The LED next to V F jack will turn on P...

Страница 10: ...yed voltage If the reading displayed does not have a suf cient number of digits i e 002 instead of 2 1011 set the function switch to the next lower range Repeat until you have maximum digits possible...

Страница 11: ...Current Measurement CAUTION Do not exceed 15 seconds when measuring the 20A range and wait for 30 minutes between each measurement Set the function switch to 20 on the DC Amps scale The LED above 20A...

Страница 12: ...OBTAINED a Less than 200mA AC Current Measurement Set the function switch to 200m on the AC AMPS scale The LED next to TEMP mA jack will turn on Plug the red test lead in TEMP mA jack and the black t...

Страница 13: ...to the meter 5 RESISTANCE MEASUREMENT WARNING NEVER CONNECT THE TEST LEAD TO ANY VOLTAGE WHEN THE FUNCTION SWITCH IS SET TO REMOVE ALL POWER FROM THE CIRCUIT BEING TESTED WHEN CHECK ING RESISTANCE DIS...

Страница 14: ...ITY TEST WARNING NEVER CONNECT THE TEST LEADS TOANY VOLTAGE WHEN THE FUNC TION SWITCH IS SET TO REMOVE ALL POWER FROM THE CIRCUIT BEING TESTED WHEN CHECKING RESISTANCE DISCHARGE ANY CHARGED CAPACITORS...

Страница 15: ...into V F jack and the black test lead into jack Apply the test leads across the diode terminals and note the meter reading Reverse the diode and note this reading Based on the reading the result can...

Страница 16: ...ocouple adapter model no TL 340 are required Set the function switch to TEMP The LED next to TEMP mA jack will turn on Observing the polarity plug in the temperature probe adapter with inserted in TEM...

Страница 17: ...I appears on the display the value of the capacitor is over the maximum range of the meter 11 HOLD FUNCTION This function is used to hold a reading When this pushbutton is pressed the data being disp...

Страница 18: ...FUSE REPLACEMENT WARNING DISCONNECT BOTH TEST LEADS FROM ANY SOURCE OF VOLTAGE BEFORE REMOVING THE BACK COVER DO NOT OPERATE THE METER UNTIL THE BACK COVER IS IN PLACE AND FASTENED SECURELY a 0 5 Amp...

Страница 19: ...open Remove the batteries Carefully lift the top of the case away from the bottom CAUTION Lift the top case carefully do not break the wires attached to the battery compartment Gently lift the battery...

Страница 20: ...couple Adapter TL 340 Test Leads with Screw on Alligator Clips TL 222 LIMITED WARRANTY Circuit Test Electronics warrants to the original purchaser that this product be free of defect in material or wo...

Отзывы: