Chapter 8 Preventive care
000190-09 NIOX VERO
®
User Manual US
33
8
Preventive care
8.1
General care
In the following sections, actions for preventive care and maintenance are de-
scribed. Do NOT try to repair the instrument. Any attempt will make the war-
ranty invalid and performance according to the specifications cannot be
guaranteed.
WARNING! DO NOT clean the instrument or handle with
products containing alcohol. This includes sprays or
wipes containing alcohol!
WARNING! DO NOT clean area immediately surrounding the
NIOX VERO
®
with products containing alcohol. This includes
sprays or wipes containing alcohol.
1.
Clean the instrument with a cloth dampened with water or a
mild
soap
solution.
CAUTION!
Minimize use of solvents
2.
Clean the breathing handle with a cloth dampened with water or a
mild
soap solution.
Note:
The breathing handle and patient filter are not intended for
sterilization.
WARNING!
•
The breathing handle and the instrument can not be cleaned with an
aerosol.
•
Do not use disinfectants or wipes containing alcohol, these might
permanently damage the sensor and instrument.
•
Do not use spray detergents.
A93
Sensor expiration date has passed
Replace the Sensor, see page 35.
A94
The breathing handle has expired
Press OK. Change breathing handle.
See page 34.
CAUTION!
The breathing handle’s NO
scrubber contains potassium
permanganate and should be disposed
of as hazardous waste in accordance
with local waste disposal regulations.
A95
Breathing handle expiration date has
passed
Replace the handle, see page 34.
It is still possible to view measurements
stored in the instrument memory.
Instrument
alerts
Screen
Action