Chapter 7 Troubleshooting
000190-09 NIOX VERO
®
User Manual US
29
A03
Unstable system
Remove any sources of disturbances
(such as cordless phones, mobile
phones or gas emitting appliances).
Wait for the system to stabilize.
A04
Count down time
The remaining time until the instrument
is ready to use is displayed.
A05
Lock MMI
When the instrument is connected to a
PC the main view buttons will be locked.
A06
Configuration code error
Only provided by Circassia upon
request.
The configuration code entered is
incorrect. Enter correct configuration
code.
If this error continues to be shown
contact Circassia Technical support.
A07
Lid open warning
Check if the battery or sensor lid is open
and close if needed. Click the OK button
when finished.
Instrument
alerts
Screen
Action
A08
Battery problem
Low power in battery or other failure.
Change the battery and click the OK
button when finished.
A09
Condensation countdown
Too frequent use of the instrument.
Remaining time until instrument has
dried out.
A15
Condensation alert
Reduce frequency of measurements.
Continue measuring at this frequency
causes condensation in the instrument
and will make the instrument unusable
for 30 minutes
A22
Memory access failure
Contact Circassia Technical support.