Chapter 7 Troubleshooting
30
000190-09 NIOX VERO
®
User Manual US
A25
Temperature or base line failed to
stabilize within 30 minutes
Check that the ambient temperature and
relative humidity is within specification. If
necessary, move the instrument to
another location and restart the
instrument.
A26
Self test failure
The self test of the instrument failed.
Restart the instrument.
If alert code persists contact Circassia
Technical support.
A27
Internal hardware error
unrecoverable
Contact Circassia Technical support.
A28
Internal hardware error recoverable
Check that the Sensor, battery and lid is
in its correct position, also make sure
that the tube is not folded
When finished restart the instrument.
Instrument
alerts
Screen
Action
A29
Analysis failure
Ambient measurement failure.
Click the OK button and obtain a new
measurement.
A30
Bluetooth connection error
Check the Bluetooth connection with the
PC.
When finished click the OK button.
A31
USB connection error
Check the USB connection with the PC.
When finished click the OK button.
A40
No Sensor inserted
Insert Sensor. See page 35
(replacement of sensor) or page 6(initial
placement of sensor).
Instrument
alerts
Screen
Action