![Circassia NIOX VERO Скачать руководство пользователя страница 16](http://html.mh-extra.com/html/circassia/niox-vero/niox-vero_user-manual_2607894016.webp)
Chapter 5 Using NIOX VERO
®
14
000190-09 NIOX VERO
®
User Manual US
The result is then displayed: (A) Patient ID - if applicable, (B) FeNO value in
ppb (parts per billion), (C) Measurement mode,
(D) Measurement sequence number, (E) OK - returns to main view.
5.4
Demonstration mode
To help professionals in guiding patients, the instrument contains three ani-
mated demonstrations with visual and audio guides of the different stages of
a measurement procedure.
1.
Select the
Animation
button on the main menu.
2.
Select which animation to use Cloud, Balloon or Meter
3.
Select the
Demo
button.
4.
Select the
forward
button to move to the following sequence.
5.
The
undo
button closes the demonstration and returns the animation
select.
6.
Select
OK
button
to confirm the changes.
7.
The
undo
button returns to the main menu without saving changes.
a.
Inhalation
through the
breathing
handle.
b.
Exhalation
through the
breathing
handle with
correct
pressure.
c.
Exhalation
through the
breathing
handle with
pressure too
weak.
d.
Exhalation
through the
breathing
handle with
pressure too
strong.