
OPERATOR MANUAL
5. Human
Machine Interface
58
5.5.11
Touch calibration [F11]
This function allows to calibrate the display LCD touch screen
From SERVICE menu, press:
or
[F11]
To come back, press:
or
(ESC)
Touch the cross that appear on the screen (in different positions)
At the end of procedure, machine restart automatically.
Calibration data are stored on the CPU.
In case of CPU replacement, it may be necessary to perform the calibration
procedure.
Содержание ME550
Страница 1: ...OPERATOR MANUAL EN ME550 DOG TAG MO ME550DT EN R4 ...
Страница 18: ...OPERATOR MANUAL 2 General information 14 ...
Страница 43: ...OPERATOR MANUAL 5 Human Machine Interface 39 ...