MetCam Technical Manual

 

Document P/N A1604870010 

11 

 

4.3.2 

Preparing the Cable End 

Cable properties to comply with MetCam terminal block: 

 

 

 

 

The overall conductor length, from the cable gland entrance and towards the terminal block, depends on where the cable 
ends inside the cable gland and the position of the conductor within the cable. As a reference, the minimal distance from 
the center of the internal collar to the edge ports of the terminal block is shown in the following: 

The straight line distance from the center of the cable gland port (the upper 
part of it) to the leftmost terminal block port is 45 mm. to this length, one must 
add the strip-length (10 mm) and some slack according to the type of 
conductors within the cable in use. 

The final overall length should add to this distance the amount in which the 
cable gland is recessed inside the collar.  

 

 

 

 

4.3.3 

Opening and Closing the Back Cover for Accessing the Terminal Block 

 

Do not open when energized! 

The cover at the back of MetCam needs to be opened for accessing the terminal block and then needs to be closed 
properly so that the unit will retain its explosion/flame proof certification. 

The arrow in the picture to the left shows the set-screw that secures the cover.  
To open the back cover, loosen the set screw using a 0.05 in. hex wrench, and 
turn the back cover counter clock-wise.  
The set-screw could end up at locations other than that shown since it is part of 
the back cover. 

A lanyard attaches the back cover to the MetCam body (on the inside) so it 
cannot drop when opened. 

Once all the electrical connections to the terminal block are done, hand-tighten 
the back cover by turning it clock-wise. Verify the back cover has reached the 
end-stop and secure it by tightening the set-screw.  
The recommended torque for the set-screw is 0.45 Nm (4 lb./in.) 

Terminal block connection technology 

CAGE CLAMP® S 

Conductor size solid 

0.2 - 1.5 mm

2

 

Conductor size fine-stranded 

0.2 - 1.5 mm

2

 

Conductor size fine-stranded 

0.25 - 1 mm2  (with insulated ferrule) 

Conductor size fine-stranded 

0.25 - 1 mm

2

  (with uninsulated ferrule) 

AWG 

24 - 16 

Strip-length 

9 - 10 mm/ 0.35 - 0.39 in 

Conductor entry angle 

0

0

 to PCB 

Содержание MetCam A160100 0 Series

Страница 1: ...MetCam Technical Manual Document P N A1604870010 1 MetCam Models A160100xxx0 Instructions for Use Manual Rev B Feb 2020...

Страница 2: ...Type Marking 6 4 Use 7 4 1 Installation Prerequisites 7 4 1 1 Electrical Installation Prerequisites 7 4 2 Installation 8 4 2 1 Handling 8 4 2 2 Installing the Tilt Mount 9 4 2 3 Connecting MetCam to t...

Страница 3: ...ocument P N A1604870010 3 5 Troubleshooting 14 6 Maintenance 14 6 1 Optical Cleaning Instructions 14 6 2 Exterior Cleaning Instructions 14 6 3 Periodic Procedures 15 7 Disposal 15 8 Technical Data 15...

Страница 4: ...ational rules and regulations associated with this product Only trained and competent personnel who are familiar with local codes and practice are permitted to inspect repair and service the product a...

Страница 5: ...d The threaded back plate shall be tightened all the way to the stopping shoulder and secured against self loosening by tightening the set screw using hex 0 050 Lens guards must always be applied and...

Страница 6: ...as a supplementary detection device for total risk reduction The limitations of this technology must be assessed with respect to their ability or inability to determine the concentration of gas There...

Страница 7: ...here is a significant likelihood of equipment personnel or moving objects accidentally knocking MetCam out of alignment should be avoided If such locations cannot be avoided measures including improve...

Страница 8: ...and wire MetCam see Section 4 3 1 MetCam power up see Section 4 4 4 2 1 Handling Warranty void if opening MetCam enclosure other than the back cover used for accessing the electrical connections to th...

Страница 9: ...mount is not symmetrical make sure to install it in the correct orientation Note that the tilt mount does not include a pan degree of freedom so it is important to align it to the monitored area when...

Страница 10: ...eld optional item should be mounted after connecting MetCam to the tilt mount It does not interfere with wiring or with aiming MetCam 4 2 5 Installing the Wi Fi Antenna MetCam can be ordered with an o...

Страница 11: ...the back of MetCam needs to be opened for accessing the terminal block and then needs to be closed properly so that the unit will retain its explosion flame proof certification The arrow in the pictur...

Страница 12: ...vailable for connecting the cable shield on the cable gland entrance collar Terminal Block Pin Assignment Pin Marking on Terminal Block Description 1 Voltage input Max 32 V 2 Voltage input 3 4 20 mA H...

Страница 13: ...cation of the Indicator LED the green arrow is pointing to it 4 5 4 20 mA Signaling and LED Status MetCam uses a stepped output and needs to be installed as a 4 wire transmitter The signaling range is...

Страница 14: ...reflector periodically The frequency of cleaning operations depends upon the local environmental conditions and specific applications It is recommended to visually inspect and clean if deemed necessa...

Страница 15: ...t be disposed of as household waste You can return this product to CI Systems free of charge For information please contact the national marketing organizations or CI Systems 8 Technical Data Mechanic...

Страница 16: ...temperature range w o wifi A1601001340 Extended temperature range wifi M25 Gland Entrance MetCam Models P N Description A1601000200 Standard temperature range w o wifi A1601000240 Standard temperature...

Отзывы: