background image

 

MetCam Technical Manual

 

Document P/N A1604870010 

 

 

Any electrical interference induced onto the conductors of a 4 

– 20 mAmp loop, by the installation, must be kept below the 

levels necessary to comply with the requirements of EN60079-29-1 or other applicable gas detector performance 
requirements / standards.  
In practice, this means that peak noise currents induced on the current loop should be no greater than ±0.25 mA. 

 

4.2

Installation 

The correct order for installing MetCam is: 

 

Install the pan/tilt mount, see 

Section 

‎4.2.2

 

 

Connect MetCam to the pan/tilt mount, see 

Sections 

‎4.2.3, ‎4.2.4 

and

 

‎4.2.5

 

 

Prepare the cable gland and wire MetCam, see 

Sectio

‎4.3.1

 

 

MetCam power-up, see 

Section 

‎4.4

 

4.2.1 

Handling 

 

Warranty void if opening MetCam enclosure, other than the back cover used for accessing the electrical 
connections to the terminal block. 

 

Do not touch MetCam optics with bare hands; take care not to contaminate the optical surfaces with any greasy 
substances. 

The three MetCam optical surfaces are numbered in the following drawing: 

– VIS window with lens guard 

– IR window 

3 - Mirror 

 

For cleaning MetCam optics use only wipes approved for optical 
equipment with 

Isopropyl Alcohol

. Failure to comply could result in 

damage. 

 

MetCam is available with M25 or 3/4" NPT ports. 

 

Make sure to use the gland or conduit connector that fits the  
MetCam port type. 

 

Using a gland with a flexible cable will allow ease of alignment 
and is therefore recommended. 

 

 

 

Содержание MetCam A160100 0 Series

Страница 1: ...MetCam Technical Manual Document P N A1604870010 1 MetCam Models A160100xxx0 Instructions for Use Manual Rev B Feb 2020...

Страница 2: ...Type Marking 6 4 Use 7 4 1 Installation Prerequisites 7 4 1 1 Electrical Installation Prerequisites 7 4 2 Installation 8 4 2 1 Handling 8 4 2 2 Installing the Tilt Mount 9 4 2 3 Connecting MetCam to t...

Страница 3: ...ocument P N A1604870010 3 5 Troubleshooting 14 6 Maintenance 14 6 1 Optical Cleaning Instructions 14 6 2 Exterior Cleaning Instructions 14 6 3 Periodic Procedures 15 7 Disposal 15 8 Technical Data 15...

Страница 4: ...ational rules and regulations associated with this product Only trained and competent personnel who are familiar with local codes and practice are permitted to inspect repair and service the product a...

Страница 5: ...d The threaded back plate shall be tightened all the way to the stopping shoulder and secured against self loosening by tightening the set screw using hex 0 050 Lens guards must always be applied and...

Страница 6: ...as a supplementary detection device for total risk reduction The limitations of this technology must be assessed with respect to their ability or inability to determine the concentration of gas There...

Страница 7: ...here is a significant likelihood of equipment personnel or moving objects accidentally knocking MetCam out of alignment should be avoided If such locations cannot be avoided measures including improve...

Страница 8: ...and wire MetCam see Section 4 3 1 MetCam power up see Section 4 4 4 2 1 Handling Warranty void if opening MetCam enclosure other than the back cover used for accessing the electrical connections to th...

Страница 9: ...mount is not symmetrical make sure to install it in the correct orientation Note that the tilt mount does not include a pan degree of freedom so it is important to align it to the monitored area when...

Страница 10: ...eld optional item should be mounted after connecting MetCam to the tilt mount It does not interfere with wiring or with aiming MetCam 4 2 5 Installing the Wi Fi Antenna MetCam can be ordered with an o...

Страница 11: ...the back of MetCam needs to be opened for accessing the terminal block and then needs to be closed properly so that the unit will retain its explosion flame proof certification The arrow in the pictur...

Страница 12: ...vailable for connecting the cable shield on the cable gland entrance collar Terminal Block Pin Assignment Pin Marking on Terminal Block Description 1 Voltage input Max 32 V 2 Voltage input 3 4 20 mA H...

Страница 13: ...cation of the Indicator LED the green arrow is pointing to it 4 5 4 20 mA Signaling and LED Status MetCam uses a stepped output and needs to be installed as a 4 wire transmitter The signaling range is...

Страница 14: ...reflector periodically The frequency of cleaning operations depends upon the local environmental conditions and specific applications It is recommended to visually inspect and clean if deemed necessa...

Страница 15: ...t be disposed of as household waste You can return this product to CI Systems free of charge For information please contact the national marketing organizations or CI Systems 8 Technical Data Mechanic...

Страница 16: ...temperature range w o wifi A1601001340 Extended temperature range wifi M25 Gland Entrance MetCam Models P N Description A1601000200 Standard temperature range w o wifi A1601000240 Standard temperature...

Отзывы: