Freeze-dryer Gamma 1-16 LSCplus
Freeze-dryer Gamma 2-16 LSCplus
5 Set-up and connection
Version 04/2013, Rev. 2.10 of 09/03/2020
• sb-dc
41 / 126
Translation of the original operating manual
• Take the vacuum sensor out of its original packaging and fasten it to the
connector with a clamping ring (DIN16KF) and a centring ring (included
in the scope of supply).
• Plug the connector to the connection socket and hand-tighten the
screws on the connector.
NOTE
It is absolutely essential to comply with the
manufacturer’s instructions in
the separate operating manual of the vacuum sensor!
Thyracont VCP 63 (Pirani)
Pfeiffer CMR 363 (capacitive)
Fig. 13: Vacuum sensors of different manufacturers
NOTE
The vacuum sensor comes supplied in a calibrated state.
After the freeze-dryer has been switched on, the vacuum sensor needs
several minutes until it is ready for operation.
5. 4. 6
Vac uum pum p
→
5.6
Vacuum pump
NOTE
It is absolutely essential to refer to the separate instruction manual of the
vacuum pump and exhaust filter (if applicable)!
The vacuum pump must be connected to the vacuum connection of the unit
and to the electrical socket at the back of the unit, which is marked
accordingly (see chapter 2.1.1 - "Functional and operating elements").