INSTRUCTION MANUAL
PULSE OXIMETER
MD300I
MD300I
- 12 -
- 13 -
CHAPTER 4 Settings
4.1 Brief Introduction of Setting
You can enter or exit the setting only under “measure mode” and “information display mode”
in order to set the brightness, patient ID, date and time.
Enter setting:
Press the function key for more than one second, the oximeter will enter
setting.
Save and exit setting:
Press function key and setting key simultaneously, the modification
under the setting mode will be saved, at the same time, the system will exit from setting mode
and return to “non-setting mode”.
Cancel and exit from setting:
Under the “setting mode”, press function board twice, the
modification under setting mode will be canceled, at the same time, the system will exit from
setting and return to “non-setting mode”.
If there is no operation under “setting mode” in 10 seconds, the modification under “setting
mode” will be canceled automatically, at the same time, the system will exit from “setting
mode” and return to “non-setting mode”.
4.2 Parameters Change
Enter the Setting Mode under the “Measure mode”, you can only set Brightness and Patient
ID. Enter the Setting Mode under the “Information Display mode”, you can set Brightness,
Patient ID, Date and Time. Current parameters and data will be displayed at the top right
corner which is used to display the “PR”. The various parameters display title is as follows:
Br(brightness)-->ID(patientID)-->Y(year)-->M(month)-->D(date)-->H(hour)-->m(minute)--
>S(second)--Br(brightness)-->...
Br: brightness, range: 1-7.
ID: patient ID, range: 1-10.
Y: year, range: 0-99, short for 2000-2099.
M: month, range: 1-12.
D: day, range: 1-31.
H: hour, range: 0-23.
m: minute, range: 0-59.
S: second, range: 0-59.
Change current value of the parameter
Press “setting key”, current parameter setting will be added by 1 unit.
Double press or longtime press “setting key”, current parameter setting will be added by 10
units.
4.3 Brightness Adjustment
Holding the function key (left key) for more than one second, the brightness level headed with
“Br” will be shown on the top right of the screen. You can adjust the brightness by degrees by
pressing the setting key (right key). There are 7 levels of brightness in all. The default is level 3.
Fig.4-1
4.4 Data Replay and Transmission
The oximeter can record SpO
2
and PR value for more than 72 hours, and can analyze
records one by one. You can transfer the history data to a PC by using “MEDVIEW”
software and an attached data cable. As for detailed setup and operation, please refer to the
“MEDVIEW” operator’s manual.