1 Introduction

1.1 Brief Introduction

Thank you for purchasing the handheld pulse oximeter for the functional oxygen saturation of 
arterial hemoglobin (SpO

2

) and Pulse Rate (PR) measurement. The pulse oximeter features 

PR tone modulation, data storage and data transmission capabilities. Please read the user 
manual carefully before using this instrument.

Note: 

• 

The illustrations applied in the manual may differ slightly from the actual device.

• 

The specifications are subject to change without prior notice. 

• 

The device is designed of handheld structure and please be sure not to turn upside 
down when using it.

1.2 Intended Use

The MD300I pulse oximeter is a portable non-invasive device intended for spot-checking of 
oxygen saturation of arterial hemoglobin (SpO

2

) and pulse rate of adult and pediatric patient 

at home and hospital (including clinical use in internist/surgery, Anesthesia, intensive care 

and etc). It is not for continuous monitoring.

1.3 Measurement Principle

Principle of the oximeter is as follows: A mathematical formula is established making use 
of Lambert Beer Law according to Spectrum Absorption Characteristics of Reductive 
hemoglobin (RHb) and Oxyhemoglobin (HbO2) in glow and near-infrared zones. Operation 
principle of the instrument: Photoelectric Oxyhemoglobin Inspection Technology is adopted 
in accordance with Capacity Pulse Scanning and Recording Technology, so that two beams 
of different wavelength of lights (660nm glow and 940nm near infrared light) can be focused 
onto a human nail tip through a clamping finger-type sensor. A measured signal obtained by a 

photosensitive element, will be shown on the oximeter’s display through process in electronic 

circuits and microprocessor.

Diagram of Operation Principle

1. Red and Infrared-ray Emission Tube
2. Red and Infrared-ray Receipt Tube

1.4 Safety Information

Conception of Warning, Caution and Note
The Warning, Caution and Note at this document are special information in favor of user’s 
operation.
• 

Warning - Indicates a potential hazard or unsafe practice that, if not avoided, will result 
in death or serious injury.

• 

Caution - Indicates a potential hazard or unsafe practice that, if not avoided, could 
result in minor personal injury or product/property damage.

Содержание MD300I

Страница 1: ...PULSE OXIMETER INSTRUCTION MANUAL MD300I...

Страница 2: ...to is protected by copyrights or patents and does not convey any license under the patent rights of our company nor the rights of others We do not assume any liability arising out of any infringements...

Страница 3: ...follows A mathematical formula is established making use of Lambert Beer Law according to Spectrum Absorption Characteristics of Reductive hemoglobin RHb and Oxyhemoglobin HbO2 in glow and near infra...

Страница 4: ...accompanying documents This equipment should not be used adjacent to or stacked with other equipment This equipment is not intended for use during patient transport outside the healthcare facility Whe...

Страница 5: ...Venous pulsations Placement of a sensor on an extremity with a blood pressure cuff arterial catheter or intravascular line The patient has hypotension severe vasoconstriction severe anemia or hypother...

Страница 6: ...PC with USB cable for data transmission Function key at the left of the oximeter acts as a power on switch when the oximeter is power off While the oximeter is power on it acts as a function key Setti...

Страница 7: ...sor cable upside when inserting the sensor 2 Do not insert the sensor forcefully CHAPTER 3 Take a Measurement Definition of key press There are three ways to press the key 1 Press press the key quickl...

Страница 8: ...l warning SpO2 and PR If the measured SpO2 and or PR value is beyond the default limit the corresponding value will flash with audible sound By default SpO2 The upper limit 100 The lower limit 90 PR T...

Страница 9: ...he Information Display mode you can set Brightness Patient ID Date and Time Current parameters and data will be displayed at the top right corner which is used to display the PR The various parameters...

Страница 10: ...Definitions Err 1 program memory damaged Err 2 data memory damaged Err 3 sensor Red Emission Diode damaged Err 4 sensor Infrared ray Emission Diode damaged Err 5 sensor Infrared ray Receipt Diode dam...

Страница 11: ...belong to consumption and beyond free change range Such damage of probe caused by mechanical force doesn t belong to free change range During measurement of SPO2 principle leads to measuring value di...

Страница 12: ...ting Temperature 5 40 Storage Temperature 20 55 Operating Humidity 80 RH no condensation Storage Humidity 93 RH no condensation Classification per IEC60601 1 Classification according to IEC 60601 1 Ac...

Страница 13: ...e model MD300I Pulse Oximeter uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment 5 RF e m...

Страница 14: ...in such an environment Immunity test IEC 60601 test level Compliance level Electromagnetic environment guidance Conducted RF IEC 61000 4 6 Radiated RF IEC 61000 4 3 3Vrms 150kHz to 80MHz 3V m 80MHz to...

Страница 15: ...r as recommended below according to the maximum output power of the communications equipment Rated maximum output of transmitter W Separation distance according to frequency of transmitter m 150KHz to...

Отзывы: