ChoiceMMed MD300I Скачать руководство пользователя страница 10

INSTRUCTION  MANUAL

PULSE  OXIMETER

MD300I

MD300I

- 14 -

- 15 -

CHAPTER 5  Maintenance and Repair

Warnings! 

• 

The advanced circuit inside the oximeter does not require periodic calibration and 
maintenance, except replacing the batteries.

• 

Don’t open the cover of oximeter or repair electronic circuits. Its open will cause the 
damage of the device and the annulment of the guarantee.

5.1 Maintenance

Replace the batteries in a timely manner when low voltage lamp is lighted.
Clean surface of the oximeter before it is used in diagnosis for patients.
Remove the batteries if the oximeter is not operated for a long time.
It is best to store the product in -20

℃~

+55

 and ≤93% humidity.

Keep in a dry place. Extreme moisture may affect oximeter lifetime and may cause damage.
Dispose of batteries properly; follow any applicable local battery disposal laws.

5.2 Cleaning and Calibrating

Please use medical alcohol to clean the silicone touching the finger inside of SpO

2

 probe with 

a soft cloth dampened with 70% isopropyl alcohol. Also clean the being tested finger using 
alcohol before and after each test.
Do not pour or spray liquids onto the oximeter, and do not allow any liquid to enter any 
openings in the device. Allow the oximeter to dry thoroughly before reuse.

A functional tester cannot be used to assess the accuracy of a pulse oximeter monitor 

or sensor. Clinical testing is used to establish the SpO

2

 accuracy. The measured arterial 

hemoglobin saturation value (SpO

2

) of the sensors is compared to arterial hemoglobin 

oxygen (SaO

2

) value, determined from blood samples with a laboratory CO-oximeter. The 

accuracy of the sensors in comparison to the CO-oximeter samples measured over the SpO

2

 

range of 70–99%. Accuracy data is calculated using the root-mean-squared (Arms value) for 
all subjects, per ISO 9919:2005, Medical Electrical Equipment–Particular requirements for the 
basic safety and essential performance of pulse oximeter equipment for medical use.

5.3 Troubleshooting

a) Error Definitions
Err 1: program memory damaged.
Err 2: data memory damaged.
Err 3: sensor

 

Red Emission Diode damaged.

Err 4: sensor Infrared-ray Emission Diode damaged. 
Err 5: sensor Infrared-ray Receipt Diode damaged.
Err 6: exterior crystal oscillator damaged.

Err 7: sensor emission diode or receipt diode damaged.
Err 9: real time clock damaged.
Err 10: EEPROM chip damaged.
b) Possible problems and corresponding solutions

Problems

Possible Reason

Solution

SpO

2

 or PR

 cannot be 
displayed 
normally

1. Finger is not plugged 

correctly

2.  Patient’s  Oxyhemoglobin 

value is too low to be measured

1. Retry by plugging the finger
2. Attempt several times to obtain 

a reading, If you are sure that no 
problem exists, obtain further clinical 
examination

SpO

2

  or  PR 

d i s p l a y   i s 
unstable

1. Finger might not be plugged 

deep enough

2. Finger is trembling or patient 

is moving continually

1. Retry by plugging the finger
2. Urge the patient to remain still

The 

Oximeter 
cannot be 
powered on

1.  Battery  power  may  be 

inadequate or not installed

2. Batteries might be installed 

incorrectly

3. The Oximeter might be 

damaged

1. Please replace batteries
2. Please reinstall the batteries
3.Contact local customer Technical 
Service

“Error3”  or  
“Error4” 
Displayed on 

screen

1.  Receiving  diode  may  be 

shielded or damaged together 

with broken connector.
2.  Mechanical  Misplace  for 
receive-emission diode
3. Amp circuit malfunction.

1.Contact local customer Technical 
Service
2.Contact local customer Technical 
Service
3.Contact local customer Technical 
Service

“Error7” 

displayed on 
screen

1. Emission diode damaged. 
2. Current control circuit 
malfunction.

1 Contact local customer Technical 
Service
2 Contact local customer Technical 
Service

“Probe  off” 

displayed on 
screen

1. The sensor is not connected
2 The connection between the 
Probe and Oximeter is loose

1.Connect the sensor
2. Please check if the probe was 

connected with oximeter correctly

5.4 Warranty and Repair

5.4.1 Service Method

Service support: Our company will offer user telephone and e-mail support as well as in-
house repair at our facility.
Parts replacement: Our company will replace parts, accessories, free of charge during the 
warranty period.

Содержание MD300I

Страница 1: ...PULSE OXIMETER INSTRUCTION MANUAL MD300I...

Страница 2: ...to is protected by copyrights or patents and does not convey any license under the patent rights of our company nor the rights of others We do not assume any liability arising out of any infringements...

Страница 3: ...follows A mathematical formula is established making use of Lambert Beer Law according to Spectrum Absorption Characteristics of Reductive hemoglobin RHb and Oxyhemoglobin HbO2 in glow and near infra...

Страница 4: ...accompanying documents This equipment should not be used adjacent to or stacked with other equipment This equipment is not intended for use during patient transport outside the healthcare facility Whe...

Страница 5: ...Venous pulsations Placement of a sensor on an extremity with a blood pressure cuff arterial catheter or intravascular line The patient has hypotension severe vasoconstriction severe anemia or hypother...

Страница 6: ...PC with USB cable for data transmission Function key at the left of the oximeter acts as a power on switch when the oximeter is power off While the oximeter is power on it acts as a function key Setti...

Страница 7: ...sor cable upside when inserting the sensor 2 Do not insert the sensor forcefully CHAPTER 3 Take a Measurement Definition of key press There are three ways to press the key 1 Press press the key quickl...

Страница 8: ...l warning SpO2 and PR If the measured SpO2 and or PR value is beyond the default limit the corresponding value will flash with audible sound By default SpO2 The upper limit 100 The lower limit 90 PR T...

Страница 9: ...he Information Display mode you can set Brightness Patient ID Date and Time Current parameters and data will be displayed at the top right corner which is used to display the PR The various parameters...

Страница 10: ...Definitions Err 1 program memory damaged Err 2 data memory damaged Err 3 sensor Red Emission Diode damaged Err 4 sensor Infrared ray Emission Diode damaged Err 5 sensor Infrared ray Receipt Diode dam...

Страница 11: ...belong to consumption and beyond free change range Such damage of probe caused by mechanical force doesn t belong to free change range During measurement of SPO2 principle leads to measuring value di...

Страница 12: ...ting Temperature 5 40 Storage Temperature 20 55 Operating Humidity 80 RH no condensation Storage Humidity 93 RH no condensation Classification per IEC60601 1 Classification according to IEC 60601 1 Ac...

Страница 13: ...e model MD300I Pulse Oximeter uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment 5 RF e m...

Страница 14: ...in such an environment Immunity test IEC 60601 test level Compliance level Electromagnetic environment guidance Conducted RF IEC 61000 4 6 Radiated RF IEC 61000 4 3 3Vrms 150kHz to 80MHz 3V m 80MHz to...

Страница 15: ...r as recommended below according to the maximum output power of the communications equipment Rated maximum output of transmitter W Separation distance according to frequency of transmitter m 150KHz to...

Отзывы: