![ChoiceMMed MD300I Скачать руководство пользователя страница 10](http://html1.mh-extra.com/html/choicemmed/md300i/md300i_instruction-manual_2602921010.webp)
INSTRUCTION MANUAL
PULSE OXIMETER
MD300I
MD300I
- 14 -
- 15 -
CHAPTER 5 Maintenance and Repair
Warnings!
•
The advanced circuit inside the oximeter does not require periodic calibration and
maintenance, except replacing the batteries.
•
Don’t open the cover of oximeter or repair electronic circuits. Its open will cause the
damage of the device and the annulment of the guarantee.
5.1 Maintenance
Replace the batteries in a timely manner when low voltage lamp is lighted.
Clean surface of the oximeter before it is used in diagnosis for patients.
Remove the batteries if the oximeter is not operated for a long time.
It is best to store the product in -20
℃~
+55
℃
and ≤93% humidity.
Keep in a dry place. Extreme moisture may affect oximeter lifetime and may cause damage.
Dispose of batteries properly; follow any applicable local battery disposal laws.
5.2 Cleaning and Calibrating
Please use medical alcohol to clean the silicone touching the finger inside of SpO
2
probe with
a soft cloth dampened with 70% isopropyl alcohol. Also clean the being tested finger using
alcohol before and after each test.
Do not pour or spray liquids onto the oximeter, and do not allow any liquid to enter any
openings in the device. Allow the oximeter to dry thoroughly before reuse.
A functional tester cannot be used to assess the accuracy of a pulse oximeter monitor
or sensor. Clinical testing is used to establish the SpO
2
accuracy. The measured arterial
hemoglobin saturation value (SpO
2
) of the sensors is compared to arterial hemoglobin
oxygen (SaO
2
) value, determined from blood samples with a laboratory CO-oximeter. The
accuracy of the sensors in comparison to the CO-oximeter samples measured over the SpO
2
range of 70–99%. Accuracy data is calculated using the root-mean-squared (Arms value) for
all subjects, per ISO 9919:2005, Medical Electrical Equipment–Particular requirements for the
basic safety and essential performance of pulse oximeter equipment for medical use.
5.3 Troubleshooting
a) Error Definitions
Err 1: program memory damaged.
Err 2: data memory damaged.
Err 3: sensor
Red Emission Diode damaged.
Err 4: sensor Infrared-ray Emission Diode damaged.
Err 5: sensor Infrared-ray Receipt Diode damaged.
Err 6: exterior crystal oscillator damaged.
Err 7: sensor emission diode or receipt diode damaged.
Err 9: real time clock damaged.
Err 10: EEPROM chip damaged.
b) Possible problems and corresponding solutions
Problems
Possible Reason
Solution
SpO
2
or PR
cannot be
displayed
normally
1. Finger is not plugged
correctly
2. Patient’s Oxyhemoglobin
value is too low to be measured
1. Retry by plugging the finger
2. Attempt several times to obtain
a reading, If you are sure that no
problem exists, obtain further clinical
examination
SpO
2
or PR
d i s p l a y i s
unstable
1. Finger might not be plugged
deep enough
2. Finger is trembling or patient
is moving continually
1. Retry by plugging the finger
2. Urge the patient to remain still
The
Oximeter
cannot be
powered on
1. Battery power may be
inadequate or not installed
2. Batteries might be installed
incorrectly
3. The Oximeter might be
damaged
1. Please replace batteries
2. Please reinstall the batteries
3.Contact local customer Technical
Service
“Error3” or
“Error4”
Displayed on
screen
1. Receiving diode may be
shielded or damaged together
with broken connector.
2. Mechanical Misplace for
receive-emission diode
3. Amp circuit malfunction.
1.Contact local customer Technical
Service
2.Contact local customer Technical
Service
3.Contact local customer Technical
Service
“Error7”
displayed on
screen
1. Emission diode damaged.
2. Current control circuit
malfunction.
1 Contact local customer Technical
Service
2 Contact local customer Technical
Service
“Probe off”
displayed on
screen
1. The sensor is not connected
2 The connection between the
Probe and Oximeter is loose
1.Connect the sensor
2. Please check if the probe was
connected with oximeter correctly
5.4 Warranty and Repair
5.4.1 Service Method
Service support: Our company will offer user telephone and e-mail support as well as in-
house repair at our facility.
Parts replacement: Our company will replace parts, accessories, free of charge during the
warranty period.
Содержание MD300I
Страница 1: ...PULSE OXIMETER INSTRUCTION MANUAL MD300I...