PULSE OXIMETER
MD300I
- 24 -
Recommended separation distances between portable and mobile RF communications
equipment and the EQUIPMENT or SYSTEM-for EQIPMENT and SYSTEMS that are not
LIFE-SUPPORTING
Recommended separation distances between portable and mobile RF communications
equipment and the MD300I Pulse Oximeter
The model MD300I Pulse Oximeter is intended for use in an electromagnetic
environment in which radiated RF disturbance are controlled. The customer or the user
of the model MD300I Pulse Oximeter can help prevent electromagnetic interference
by maintaining a minimum distance between portable and mobile RF communication
equipment ( transmitters) and the model MD300I Pulse Oximeter as recommended
below, according to the maximum output power of the communications equipment.
Rated maximum
output of
transmitter(W)
Separation distance according to frequency of transmitter (m)
150KHz to 80 MHz
80MHz to 800 MHz
800MHz to 2.5 GHz
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
For transmitters rated at a maximum output power not listed above, the recommended
separation distance in meters (m) can be estimated using the equation applicable to
the frequency of the transmitter, where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter manufacturer.
NOTE1 At 80MHz and 800MHz, the separation distance for the higher frequency range
applies.
NOTE2 These guidelines may not apply in all situations. Electromagnetic interference is
affected by absorption and reflection from structures, objects and people.