Operation

 

 
● Install the supplied application software to your personal computer. 

(For the details, refer to the instruction manual for the application software.) 

 
● For more accurate measurements, wait 30 minutes after power-on until this product is stabilized. 
 
● Use this product with noiseless and stable power source. 
 
 
 

 Monitor lamp

 

 
The state of product is indicated by 3 colors on the LED indicator; blue, purple and red. 
 

Item 

Color 

Specification 

Mode is not set (Factory Default) 

Blue 

Lit once for 3 seconds 

Alarm values are not set 

Blue 

Lit once for 3 seconds 

Alarm being monitored 

Blue 

Lit continuously 

Set to Captured Mode 

Purple 

Lit continuously 

Reboot  is  required  after  all  settings  have 
completed 

Purple 

Lit for 3 seconds twice 

Alarm is activated 

Red 

Lit continuously 

Abnormal internal memory 

Red 

Lit once for 3 seconds 

* Refer to Page 14 Troubleshooting. 

Abnormal 
State 

Abnormal internal temperature 

Red 

Lit once for 3 seconds 

* Refer to Page 14 Troubleshooting. 

 
 
 

Operation mode

 

 
1. Monitor mode: When CHINO application software is used 

The  temperature  data  is  sent  continuously  to  a  master  unit  (e.g.  PC)  by  a  command  from  the                 
software.(The frame rate of the thermal image on the software is 3fps without any Alarm activated.) 

The  alarm  set  on  the  sensor  can  provide  a  non-voltage  contact  output  while  sending  the 
temperature data.(The frame rate goes down to 1fps or less with the alarm set on the sensor.) 

 

 

 

2. Capture mode: Without CHINO application software 

The temperature data in a single row of    the thermopile array can be obtained at a time    by a 
command from a master unit.   

 

     

When Capture mode is used, a series of communication commands are required. 

     

For the communication commands, contact to your nearest distributor. 

 
 
 
 
 

Содержание TP-L0260EN

Страница 1: ... store this instruction manual in a designated place for future reference INST No TPL 02 5A TP TP TP TP L L L L0260 0260 0260 0260EN EN EN EN Instruction Manual CHINO INSTRUCTIONS Compact Thermal Image Sensor ...

Страница 2: ......

Страница 3: ...structions In addition please retain this instruction manual for future reference Before using Please confirm the contents of packing If something is missing please contact to your nearest distributor Thermal image sensor Carl plugs for mounting the universal head into concrete 3 pieces Connector cap 1 piece Lens cap 1 piece Universal head 1 piece Screws for the universal head 3 pieces Mounting sc...

Страница 4: ...s instruction manual may be emphasized simplified or omitted Every effort has been made to ensure that the details of this manual are accurate However should any errors be found or important information be omitted please inform your nearest distributor Microsoft and Windows are either trademarks or registered trademarks of Microsoft Corporation USA The company names and brand names used in this ma...

Страница 5: ...d operation of this product will not be guaranteed It may also cause fire electric shock Avoid the use of this product in the following places A place where ambient temperature varies widely A place where humidity is high A place where rain water is splashed A place where there are dusts like sand or dirt A place where it is subjected to exposed in scorching heat or direct sunlight A place where i...

Страница 6: ...may cause malfunction or trouble on this product Keep this product away from equipment generating strong high frequency or surge voltage or it may cause malfunction or trouble on this product Use neutral detergent for cleaning of this product and wipe it off lightly by rubbing with the cloth squeezed firmly Do not use chemicals or bleaches including benzene thinner and alcohol Do not use or place ...

Страница 7: ...1 5 Basic configuration 2 Names and functions 4 Connection 5 Installation 7 Operation 8 Monitor lamp 8 Operation mode 8 1 Monitor mode 8 2 Capture mode 8 Alarm output 9 About reset switch 11 Specifications 12 Precautions For maintaining measurement accuracy 14 Troubleshooting 14 External dimensions 1 Thermal image sensor 15 2 Universal head 15 3 Protective case 16 ...

Страница 8: ...oltage contact outputs The sensor can be used as a fixed mount type alarm unit When an alarm is activated 1 image can be stored in the sensor 3 Functions Alarm can be set from the application software or custom made commands The sensor provides non voltage contact outputs when the alarm condition is met Monitor mode when the supplied application software is used The temperature data of the entire ...

Страница 9: ...HUB without automatic crossing straight converting function is used a straight cable is needed Connect a crossing straight converting connector between the custom Network cable and a HUB Refer to Page 3 Alarm setup or monitor with the PC Alarm contact outputs from the thermal image sensor Alarm contact outputs from the thermal image sensor Thermal image sensor PC Thermal image sensor DC power supp...

Страница 10: ...r nearest distributor Custom Network cable Crossing cable HUB Crossing straight converting connector LAN cable Straight cable PC Alarm contact output from the thermal image sensor For image confirmation and alarm settings only LAN Contact output Contact output Other alarming devices PLC Capture mode when the supplied CHINO application software is not used The temperature data in a single row of th...

Страница 11: ...age 8 2 Lens For forming image of radiation energy from a measurement object on the detecting element 3 Mounting screw For mounting the sensor on a tripod 1 4 20UNC 4 Power alarm output connector For connecting custom power alarm output cable 5 I F connector For connection of custom Network cable 6 Reset switch To initialize the sensor to factory default The switch is located behind the mounting s...

Страница 12: ... this product The connections are a quick disconnect locking type Insert the plug until it clicks To unplug hold the sliding part that has the Arrow marking on and pull it outward If the A part is pulled and the plug may be damaged Plug the connector of the custom Network cable to the Ethernet port of a master unit PC etc and connect the O tip terminals of the custom power alarm output cable to th...

Страница 13: ...400V Peak AC Max continuous load current 0 1A Peak AC Peak load current 240mA 100ms 1 shot Output capacity Per 1 contact Max output loss 300mW If the custom Network cable is not used When using the thermal image sensor stand alone Cover the supplied connector cap to the connector marked with LAN back of this product If you don t cover the connector with the cap the waterproofing performance cannot...

Страница 14: ... center of screen as large as possible Relation between a measuring surface and a measuring distance D View width 1 4 L Measuring distance For accurate measurements 8 x 8 pixels or more for a target is recomended 1 4m Do not install the sensor where subjected to vibration shock or inductive ambient 1m 3m 5m 29mm 4 2m 88mm 146mm 7 0m View per pixel ...

Страница 15: ... continuously Abnormal internal memory Red Lit once for 3 seconds Refer to Page 14 Troubleshooting Abnormal State Abnormal internal temperature Red Lit once for 3 seconds Refer to Page 14 Troubleshooting Operation mode 1 Monitor mode When CHINO application software is used The temperature data is sent continuously to a master unit e g PC by a command from the software The frame rate of the thermal...

Страница 16: ...contact output Number of outputs 2 points Alarm name AL1 alarm 1 AL2 alarm 2 Output logic Output logic selectable N O N C Alarm condition Selectable high limit low limit Alarm condition when power is off Opened 2 Image storage The image data can be stored when alarm is activated Item Content Number of storage 1 image An image can be stored by AL1 or AL2 to be specified Storage medium Built in EEPR...

Страница 17: ...コマンド コマンド コマンド コマンドまで まで まで まで受信 受信 受信 受信まで まで まで まで 保持 保持 保持 保持する する する する 自動復帰 自動復帰 自動復帰 自動復帰 保持 保持 保持 保持 Alarm output and storing image 熱画像 熱画像 熱画像 熱画像センサ センサ センサ センサの の の の電源再投入又 電源再投入又 電源再投入又 電源再投入又は は は は アプリケーションソフト アプリケーションソフト アプリケーションソフト アプリケーションソフトからの からの からの からの 警報解除 警報解除 警報解除 警報解除まで まで まで まで保持 保持 保持 保持する する する する Alarm state Auto Reset Contact output Image saving ON OFF ON Image is ove...

Страница 18: ... values becomes effective The violet monitor lamp flashes 2 times By re power on the set values are reset to the default values at the shipment from factory Usually do not push the reset switch Factory Default settings Items Specifications IP address 192 168 1 254 Subnet mask 255 255 255 0 Gateway address 0 0 0 0 Invalidation Sensor number 1 UDP port address 50828 fixed 1 Emissivity 1 00 all pixel...

Страница 19: ...ondensation Storage temperature range 20 C to 60 C Housing IP65 with custom cable connection or connector cap and mounting screw mounted External dimensions W62 x H62 x D59 mounting screw not included Weight About 150g sensor unit Material Polycarbonate Resin Color Black Conforming standard Influence within 3 of measurement range under the environment of CE EN61326 1 Class B mounting screw size 1 ...

Страница 20: ...m setting information etc DHCP Not supported Thermal image sensor IP address 192 168 1 254 default at shipment Thermal image sensor subnet mask 255 255 255 0 Default gateway 0 or FFFF Not set Connection Custom connector connection Connector name I F Accessory sold separately Protective case Model TP ZCC1 Refer to Extrernal Dimension Minimum requirements for the CHINO application software Compatibl...

Страница 21: ... opaque object Troubleshooting If you are experiencing difficulties with the sensor or noticing an abnormality on its operation refer to the followings If the difficulties or abnormality is still present contact to your nearest distributor Symptoms Causes Measures The custom power cable is not connected properly Connect the custom power cable properly Indicator lamp is not lit The power for equipm...

Страница 22: ...acket with 3 pieces of screws through the 3 holes on the bracket If the universal head needs to be mounted on concrete surface drill holes with 6 5mmø and 25mm depth on the concrete and push the Carl plugs into the holes and then mount the universal head Carl plugs For mounting on concrete The plugs are not used for usual installation Screw size 1 4 20UNC mm mm mm mm mm ...

Страница 23: ...16 3 Accessory Protective case Model TP ZCC1 Hole for air purge PT1 8 ...

Отзывы: