11.2.3 Status display items
Screen
Display
Inside temperature
Dtemp
Model
Model
Serial number
SN
Measuring temperature range
R
Measuring wavelength
WL
11.2.4. System programming display items
Screen
Display
Programming range
Default
Temperature unit
Unit
o
C,
o
F
o
C
Output scaling high limit *1
OH
-50 to 1000
o
C (IR-AHT)
600 to 3000
o
C (IR-AHS)
900 to 3000
o
C (IR-AHU)
1000
o
C (IR-AHT)
3000
o
C (IR-AHS)
3000
o
C (IR-AHU)
Output scaling low limit *1
OL
-50 to 1000
o
C (IR-AHT)
600 to 3000
o
C (IR-AHS)
900 to 3000
o
C (IR-AHU)
-50
o
C (IR-AHT)
600
o
C (IR-AHS)
900
o
C (IR-AHU)
Zero/span calibration execution
CMP
Calibration data programming at
zero side *2
Zero
-50 to 1000
o
C (IR-AHT)
600 to 3000
o
C (IR-AHS)
900 to 3000
o
C (IR-AHU)
-50
o
C (IR-AHT)
600
o
C (IR-AHS)
900
o
C (IR-AHU)
Calibration data programming at
span side *2
Span
-50 to 1000
o
C (IR-AHT)
600 to 3000
o
C (IR-AHS)
900 to 3000
o
C (IR-AHU)
1000
o
C (IR-AHT)
3000
o
C (IR-AHS)
3000
o
C (IR-AHU)
Calibration operation execution
*2
CAL
Calibration data recovery *2
Undo
*1: The output scaling high limit and output scaling low limit are only displayed on the thermometers
with analog output (option).
*2: The screens for the zero side calibration data programming, the span side calibration data
programming, the calibration operation execution and the calibration data recovery are only
displayed when you select “GO” on the zero/span calibration execution screen.