3.1 UV254 Sensor
51
Options
→
Setup
→
Modbus Setup
Slave Address
Enter the Modbus slave address of the sensor. This is unlikely to change from 1.
Mode
Select the Modbus serial mode. Options are “ASCII” or “RTU”.
Baud Rate
Select the RS485 baud rate. Options are standard baud rates between 1200 and 115200.
Parity
Select the RS485 parity bit option. Options are “Odd”, “Even”, or “None”
Stop Bits
Select the number of stop bits. Options are 1 or 2.
Termination Enabled
Use the checkbox to enable the use of a termination resistor.
Byte Swap
To support different microprocessors that store data in different formats, the HydroAct 4
can alter the data it sends and receives to match the format expected by the sensor. The byte
swap option allows the order of bytes in the 16-bit registers to be reversed.
Word Swap
When two 16-bit registers are combined for larger data types, e.g., floating-point number
representations, this option reverses the order of the two 16-bit words to match the format
expected by the sensor.
3.1.4.8
COD
Options
→
Virtual Setup
→
COD
COD Enabled
Options are “Yes” or “No”.
Correlation Parameter
Select which parameter to utilize for the linear relationship. Options are
measured UVA or UVA compensated for sample turbidity.
Slope
Enter the calibration slope calculated from the linear correlation.
Offset
Enter the calibration offset calculated from the linear correlation.
3.1.4.9
TOC
Options
→
Virtual Setup
→
TOC
Содержание HydroAct HA4
Страница 1: ...HydroAct 4 User Manual Chemtrac Inc Firmware Version 1 19 November 15 2017...
Страница 8: ......
Страница 9: ...I 2 Overview 11 2 1 Safety Precautions 2 2 Technical Data 2 3 Installation 2 4 Operation Analyzer...
Страница 10: ......
Страница 32: ......
Страница 34: ......
Страница 42: ...42 Chapter 3 Sensors For safety and correct operation the sensor must be properly earthed...
Страница 74: ......
Страница 92: ......
Страница 93: ...III 6 Warranty 95 7 Returns 97 7 1 Contact Details Documentation...
Страница 94: ......
Страница 96: ......