![Chemtrac HydroAct 4 Скачать руководство пользователя страница 37](http://html1.mh-extra.com/html/chemtrac/hydroact-4/hydroact-4_user-manual_2595953037.webp)
3.1 PC6 Particle Counter
37
The amount of light blocked from the detector is the sum of the light absorbed by the particle and
the light scattered or reflected by the particle. The size, composition, color, refractive index, light
scattering coefficient, etc of each particle will determine how much light is scattered and how much is
absorbed. The orientation of the particle as it passes through the light beam will also affect how much
light is blocked. These factors need to be taken into consideration when determining how data from
the particle counter should be interpreted. It is important to keep in mind that there are no absolutes
with any type of optical measurement that is done with light scatter or light blocking techniques. With
that said, having size data which is relatively indicative of the actual particle sizes in the sample is very
useful in categorizing the performance of a filter or determining how much relative volume of particulate
contamination exist in a given sample. The sensor output is fed into the counter electronics where the
particles are counted and sorted into size channels according to their pulse amplitude. These pulses
are compared to a calibration curve that was generated using PSL Spheres and this is how the size of
the particles is determined.
3.1.2.1
Parts Supplied
Keep the packaging for the sensor. In case of repair or warranty return the sensor in this package.
Check that the delivery is intact. In case of damage please contact your supplier. Check that the
delivery is complete by comparing with the list of parts below.
• Flow control weir
• 2 Weir clamps
• 1 Sample drain cup
• Overflow tubing
• Weir inlet & drain cup tubing
• Sensor tubing
• 400 Micron in-line strainer (Optional: Recommended for Raw Water)
• Cleaning brush
• Optional communications cable
Содержание HydroAct 4
Страница 1: ...HydroAct 4 User Manual Chemtrac Inc Firmware Version 1 19 November 15 2017...
Страница 8: ......
Страница 9: ...I 2 Overview 11 2 1 Safety Precautions 2 2 Technical Data 2 3 Installation 2 4 Operation Analyzer...
Страница 10: ......
Страница 32: ......
Страница 33: ...II 3 Sensors 35 3 1 PC6 Particle Counter 4 Outputs 55 4 1 Analog Outputs 4 2 Relay Output Options...
Страница 34: ......
Страница 54: ......
Страница 72: ......
Страница 73: ...III 5 Warranty 75 6 Returns 77 6 1 Contact Details Documentation...
Страница 74: ......
Страница 76: ......