20
Figure 4.20: HV DC programmable test menu
Test results / sub-results
U ................... measured test voltage
I .................... test current
Ic .................. capacitive portion of test current
Ir ................... resistive portion of test current
Test parameters
Starting DC test voltage
U start [500 V ... 6000 V in steps of 50 V]
DC test voltage
U test [500 V ... 6000 V in steps of 50 V]
Duration of starting voltage
t start [1 s ... 120 s ]
Duration of ramp
t ramp [2 s ... 60 s ]
Duration of test voltage
t end [Off, 1 s ... 120 s ]
Test limits
High limit (I)
H limit [0.05 mA ... 10.0 mA ]
Low limit (I)
L limit [Off, 0.05 mA ... 10.0 mA]
Test circuit
Figure 4.21: HV DC programmable test