Figure 4.38: Examples of Ipe Leakage measurement results
4.1.11 Touch Leakage
Figure 4.39: Touch Leakage test menu
Test results / sub-results
Itou ............... Touch Leakage current
P ................... Power
Test parameters
Duration
Duration [Off, 2 s ... 180 s]
Change status
Change [YES, NO]
YES: The instrument measures leakage current in two sequential steps with delay* in
between. The phase voltage is firstly applied to the right live output of the mains
test socket and secondly to the left live output of the mains test socket.
NO: The phase voltage is applied only to the right live output of the mains test socket.
*Delay time
Delay [0.2 s … 5 s]
Output connections
[Socket L,N – PE,P/S]
Test limits
H Limit (Itou)
H limit [Off, 0.25 mA ... 15.0 mA ]
L Limit (Itou)
L limit [Off, 0.25 mA ... 15.0 mA ]