9
In pre-test a too high external voltage was detected between
ISO/SUB and PE terminals. The measurement was cancelled. Press
OK
to continue.
In pre-test a possible high leakage current was detected. It is likely
that a dangerous leakage current (higher than 3.5 mA) will flow after
applying power to the device under test.
Select
YES
to proceed with or
NO
to cancel measurement.
The measured leakage (Idiff, Ipe, Itouch) current was higher than 20
mA. Measurement was aborted. Press
OK
to continue.
The load current exceeded the highest upper limit of 10 A for the
Discharging time test. Measurement was aborted. Press
OK
to
continue.
The load current continuously exceeded 10 A for more than 4 min
(moving average) in Power and Leakage tests. Measurement was
stopped for safety. Press
OK
to continue.
The load current exceeded the highest upper limit of 16 A for the
Power and Leakage tests. Measurement was aborted. Press
OK
to
continue.
Warning for restart of the instrument to set new Ethernet settings.
This message appears on exit from Settings menu after changing
Ethernet settings. Press
OK
to continue.
The instrument is overheated. The measurement can’t be carried out
until the icon disappears. Press
OK
to continue.
The device under test should be switched on (to ensure that the
complete circuit is tested).
Test voltage in Insulation resistance measurement is too low.
Measurement result is scaled to 110 V.