4.2.10 DC currents (10A DC)
(*)
Admissible overload: 10 A to 20 A for 30s max. with a 5 min pause between 2 measurements. Ambient temp. 35° C max.
4.2.11 AC currents (10A AC)
Reminder: the measurement uncertainty is specified for a display of between 550 and 6000 points. In the following tables,
refer to the specified measurement range in which uncertainty is guaranteed.
Range
Resolution
Specified
measurement
range
Uncertainty (±)
Crest
factor
Voltage
drop
Protection
6 A
0.001 A
0.6 A to 6 A
40 Hz to 1 kHz
1.2 % R + 5 D
2.8 at 5 A
10 A *
0.01 A
1 A to 10 A
40 Hz to 1 kHz
1 % R + 3 D
3.7 at 8 A
0.05 V / mA
Fast fuse
11 A/1000 V
(*) Admissible overload: 10 A to 20 A for 30 s. max. with a pause of 5 min. between 2 meas. Ambient Temp. 35° C max.
4.2.12 Max/Min
Add 0.2 % + 2 D to the uncertainty of the function and of the range used to obtain the corresponding global uncertainty.
Extrema Capture time : 100 ms.
Range
Resolution
Specified measurement
range
Uncertainty (±)
Voltage
drop
Protection
6 A
0.001 A
0.020 - 6.000 A
0.8 % R + 3 D
10 A / 20 A
*
0.01 A
0.200 - 20.00 A
0.8 % R + 2 D
0.05 V / A
Fast fuse
11 A/1000 V