19
19
4. SPECIFICATIONS
4.1 Reference conditions
Influencing quantities:
Reference conditions
Temperature
23° C ± 5° C
Relative humidity
45 % to 75 %
Power supply voltage
9 V ± 1 V
Frequency domain of the applied signal
40 Hz to 1 kHz
Absence of electric field
4.2 Specifications of the reference conditions
Uncertainties are given in:
X % of the reading (R) ± Y counts (D).
When the frequency exceeds 1 kHz, apply the formula given in the tables
X % R + Y % x [F (kHz) - 0.4] R ± D.
where:
- Reading
"R",
- Digit
"D", the measurement count equivalent to the resolution of the display range,
- Frequency
"F" in kHz.
4.2.1 DC voltages
V DC
Range
Resolution
Uncertainty (
±
)
Input resistance
600 mV
0.1 mV
0.5 % R + 2 D
6 V
0.001 V
60 V (*)
0.01 V
600 V
0.1 V
1000 V
1 V
0.2 % R + 2 D
10 M
Ω
(*) Input impedance
≈
10 M
Ω
// 50 pF