Triton
®
DTS TRT-600 Spine Therapy Table
10
SpECiFiCATiONS
ELECTrOmAgNETiC COmpATiBiLiTy TABLES
The following tables apply to the 230 Volt a/c Spine Therapy Table.
Guidance and manufacturer’s declaration – electromagnetic emissions
The Triton DTS TRT-600 Spine Therapy Table is intended for use in the electromagnetic environment
specified below. The customer or the user of the Triton DTS TRT-600 Spine Therapy Table should assure that
it is used in such an environment.
Emission tests
Compliance
Electromagnetic environment -
guidance
Harmonic emissions
IEC 61000-3-2
Class B
The Triton DTS TRT-600 Spine
Therapy Table is suitable for use in
all establishments, including
domestic establishments and
those directly connected to the
public low-voltage power supply
network that supplies buildings
used for domestic purposes.
Voltage fluctuations
IEC 61000-3-3
Complies
The Triton DTS TRT-600 Spine
Therapy Table is suitable for use in
all establishments, including
domestic establishments and
those directly connected to the
public low-voltage power supply
network that supplies buildings
used for domestic purposes.
Guidance and manufacturer’s declaration – electromagnetic immunity
The Triton DTS TRT-600 Spine Therapy Table is intended for use in the electromagnetic environment
specified below. The customer or the user of the Triton DTS TRT-600 Spine Therapy Table should assure that
it is used in such an environment.
Immunity test
IEC 60601
test level
Compliance level Electromagnetic environment -
guidance
Portable and mobile RF
communications equipment
should be used no closer to any
part of the Triton DTS TRT-600
Spine Therapy Table, including
cables, than the recommended
separation distance calculated
from the equation applicable to
the frequency of the transmitter.
Recommended separation
distance
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz to 80 MHz
Not Applicable
d
= [3.5]
√
P
V1
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz to 2.5 GHz
Not Applicable
d
= [3.5]
√
P
80 MHz to 800 MHz
E
1
d
= [7]
√
P
800 MHz to 2.5 GHz
E
1
where
P
is the maximum output
power rating of the transmitter in
watts (W) according to the
transmitter manufacturer and
d
is
the recommended separation
distance in meters (m).
Field strengths from fixed RF
transmitters, as determined by
an electromagnetic site survey
a
,
should be less than the
compliance level in each
frequency range
b
.
Interference may occur in the
vicinity of equipment marked
with the following symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and
land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted
theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an
electromagnetic site survey should be considered. If the measured field strength in the location in which the
Triton DTS TRT-600 Spine Therapy Table is used exceeds the applicable RF compliance level above, the Triton
DTS TRT-600 Spine Therapy Table should be observed to verify normal operation. If abnormal performance is
observed, additional measures may be necessary, such as reorienting or relocating the Triton DTS TRT-600
Spine Therapy Table.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.