CeFAR tens
9.3 others
Problem
Possible Cause
Solution
Display does not come on
Low battery
Charge battery
Weak stimulation with loaded
batteries
Electrodes dried out, lost their
adhesive power and have no
adequate connection to the
skin.
Replace electrode
Acupuncture mode is wrongly
selected
Unselect acupuncture mode
Electrode placement
Make the electrodes at least 5
cm (2 inches) apart
stimulation stops with loaded
batteries
Poor electrode contact
Reapply electrodes, secure
firmly. Electrodes must be a
minimum of 5 cm apart.
Damaged or worn electrodes
Replace
Cabel or contacts damaged
Replace
stimulation weakens within
minutes of starting treatment
with loaded batteries
This is a normal body adaptive
process
Increase the amplitude
(intensity) if required.
stimulation is uncomfortable
Amplitude (intensity) is too
high
Decrease amplitude (intensity).
Electrodes are too close
together
Reposition the electrodes.
Electrodes must be a minimum
of 5 cm (2 inches) apart.
Damaged or worn electrodes
Replace
Ensure proper program is being
used
A. Refer to the programs
description
B. Contact clinician if discomfort
persists