Declaration - electromagnetic emission
Cefar TENS is intended for use in the electromagnetic environment specified below. The customer or
the user of Cefar TENS should assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment -
guidance
RF emissions
CISPR 11
Group 1
Cefar TENS uses RF energy only for
its internal function. Therefore, its
RF emissions are very low and are
not likely to cause any interference
in nearby electronic equipment.
RF emissions
CISPR 11
Class B
Cefar TENS is suitable for use in all
establishments, including domestic
establishments and those directly
connected to the public low-
voltage power supply network
that supplies buildings used for
domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complies
Declaration - electromagnetic immunity
Cefar TENS is intended for use in the electromagnetic environment specified below. The customer or
the user of Cefar TENS should assure that it is used in such an environment.
Immunity Test
IEC 60601 test level
Compliance level
Electromagnetic environment
- guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±8 kV contact
±2 kV, ±4 kV, ±8 kV,
±15 kV air
±8 kV contact
±2 kV, ±4 kV, ±8 kV,
±15 kV air
Floors should be wood, concrete
or ceramic tie. If floors are
covered with synthetic material,
the relative humidity should be
at least 30 %.