284
ES
CEFAR TENS
284
CEFAR TENS
declaration - electromagnetic immunity
Cefar TENS is intended for use in the electromagnetic environment specified below. The customer or
the user of Cefar TENS should assure that it is used in such an environment.
Immunity Test
IEC 60601 test
level
Compliance level
Electromagnetic environment -
Guidance
Conducted RF
IEC 61000-4-6
3 V
0.15 MHz to 80
MHz
6 V in ISM and
amateur radio
bands between
0.15 MHz and 80
MHz
3 V
0.15 MHz to 80
MHz
6 V in ISM and
amateur radio
bands between
0.15 MHz and 80
MHz
Portable and mobile RF
communications equipment should
be used no closer to any part of
Cefar TENS, than the recommended
separation distance calculated
from the equation applicable to
the frequency of the transmitter.
Recommended separation distance
d= 3.5 √P
150 KHz to 80 MHz
V1
d= 3.5 √P
80 MHz to 800 MHz
E1
d= 7 √P
= 80 MHz to 2.7 GHz
E1
where P is the maximum output
power rating of the transmitter
in watts (W) according to the
transmitter manufacturer and d is the
recommended separation distance in
meters (m).
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey, a should
be less than the compliance level in
each frequency range.
Interference may occur in the vicinity
of equipment marked with the
following symbol:
20xx
LATEX
FREE
REF