50
Vectra Genisys® Laser Module
TROUBLESHOOTING
CODE
NUMBER
TYPE
MESSAGE
PROBABLE CAUSES
POSSIBLE REMEDIES
133
Warning
Laser Applicator became unplugged
while performing a laser treatment
A. Connect desired Laser Applicator to the system.
B. If Laser Applicator is connected, reset Therapy System by turning
main power switch Off and On.
C. Connect to a calibrated Laser Applicator.
D. Send Laser Module to Factory for Service.
134
Warning
Entered incorrect laser PIN.
Enter correct Laser PIN number.
139
Warning
Laser Board Software upgrade
warning.
Contact Chattanooga Group Service Department.
142
Warning
A User Protocol was selected but no
Laser Module is installed on system.
Install Laser Module to Therapy System. Refer to Laser User Manual (
for installation Instructions
143
Warning
A User Protocol was selected but
no Laser Applicator connected to
system.
A. Connect proper Laser Applicator to the system.
B. If Laser Applicator is connected, reset Therapy System by turning main power
switch Off and On.
C. Connect a known good Laser Applicator.
D. Send Laser Module to Factory for Service.
ERROR MESSAGES CONTINUED