49
Vectra Genisys® Laser Module
TABLE OF CONTENTS
Vectra Genisys® Laser Module
TROUBLESHOOTING
ERROR MESSAGES
CODE
NUMBER
TYPE
MESSAGE
PROBABLE CAUSES
POSSIBLE REMEDIES
130
Warning
Another treatment is running while
attempting to set up and perform
treatment.
A. Allow existing treatment to complete before starting therapy.
B. If no other treatment is running, reset Therapy System by turning main power
switch Off and On.
131
Warning
Treatment Room Door Lockout is
breached.
A. Make certain Treatment Room Door is completely closed.
B. Make certain the Laser Interlock cable is connected to the system.
C. Replace Laser Interlock to System cable with a known good cable.
D. Contact department responsible for installation of the Treatment
Room Door Laser Interlock Kit for maintenance or repair.
E. Send Laser Module to Factory for Service.
132
Warning
Attempted to start a laser treatment
but no laser applicator is plugged in.
A. Connect desired Laser Applicator to the system.
B. If Applicator is connected, reset Therapy System by turning main power switch
Off and On.
C. Connect a known good Laser Applicator.
D. Send Laser Module to Factory for Service.