27
Vectra Genisys® Laser Module
TABLE OF CONTENTS
Vectra Genisys® Laser Module
SETUP
ATTACHING THE MODULE TO THE UNIT CONTINUED
19.
Replace all cables and wires on the original unit.
20.
Insert the patient interrupt switch, laser interlock (see
), and laser applicator in the
proper connection. For more information about the symbols used for the connections,
21.
Turn the unit on using the On/Off Switch.
The system will automatically recognize the added module and display a configuration
change
message.
22.
Read and carefully follow the instructions on the screen.
WARNING
Verify that the module installed is the module displayed in the message BEFORE pressing the START button.
If it is not, DO NOT press the START button. Turn the system Off and back On. If the problem persists, call the
dealer or Chattanooga Group Technical Support immediately. DO NOT USE THE SYSTEM until all necessary
repairs are made by a technician certified by Chattanooga Group. If use is attempted before repairs are made,
the system may operate unpredictably and has the potential of causing injury to the patient or damage to
the system’s internal components.