V
Viissuuaall IInnssppeeccttiioonn
A
A.. G
Geenneerraall
Visually inspect the Fluido DHT unit. A visual
inspection can, to an experienced Technician,
indicate possible abuse of the unit and/or internal
problems.
G
Grroouunndd R
Reessiissttaannccee TTeesstt
A
A.. V
Voollttaaggee S
Sppeecciiffiiccaattiioonnss
M
Mooddeell 11448800 . . . . . Input: 120 VAC~50/60 Hz, 40 Watts
M
Mooddeell 11448811 . . . . Input: 230 VAC~50/60 Hz, 40 Watts
B
B.. S
Sppeecciiffiiccaattiioonn
Maximum Acceptable Resistance: 500 milliohms
C
C.. EEqquuiippm
meenntt R
Reeqquuiirreedd
Milliohm Meter
D
D.. TTeesstt
Place unit on level work surface.
Place one meter probe on the ground prong of power
cord and the other to any exposed metal or screw on
the unit.
LLeeaakkaaggee TTeessttss
TTeesstt V
Voollttaaggee S
Sppeecc . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1000V
Conduct all necessary leakage tests as required per
“Chapter 7 Electrical Equipment” of the 1999, or later,
edition of the NFPA (National Fire Protection
Association) “Health Care Facilities” standards.
R
Reesseettttiinngg FFlluuiiddoo D
DH
HTT FFaaccttoorryy D
Deeffaauulltt S
Seettttiinnggss
To reset all factory default settings of the unit, turn unit
Off. Depress and hold the “Clear” button on the control
panel and turn unit on simultaneously.
PPrreehheeaatt TTeesstt
A
A.. TToooollss &
& EEqquuiippm
meenntt R
Reeqquuiirreedd
Calibrated Thermometer
B
B.. PPrreehheeaatt TTeesstt PPrroocceedduurreess
Power Requirements:
Model 1480. . . . . . . . . . . . . . . . . . . . . . 120 VAC
Model 1481. . . . . . . . . . . . . . . . . . . . . . 230 VAC
Plug the unit power cord
to a grounded electrical
service receptacle conforming to the applicable
national and local electrical codes.
Turn unit power switch On.
Start the preheat function by depressing the
“PREHEAT STANDBY” button on the control panel.
After approximately 30 seconds, the Blue LED
should illuminate.
After approximately 30 minutes or less, depending
on the ambient temperature of the medium, the
blower should reduce from 50% airspeed to 5%
airspeed and the Temp indicator should read the
temperature of the Cellex
®
medium. Insert the
calibrated thermometer into medium and record the
reading. The Temperature reading should be ± 5 °F
or 3 °C of Spec below.
Preheat Default Temp. Spec
. . . . 115 °F (46.11 °C)
A temperature reading outside the specified range
may be caused by clogged intake filters, clogged
distributor, faulty temperature sensor, or faulty
heating element. Refer to the appropriate section for
the proper removal and replacement procedures.
Fluido DHT
™
Dry Heat Therapy Unit
TROUBLESHOOTING
11
!
UNIT FAILING DIELECTRIC WITHSTAND AND/OR LEAKAGE
TESTS COULD INDICATE SERIOUS INTERNAL SYSTEM
PROBLEMS.
DO NOT PLACE UNIT BACK INTO SERVICE! SEND UNIT TO
FACTORY FOR REPAIR!
DO NOT ATTEMPT TO REPAIR IN THE FIELD!
POWER
SWITCH
WARNING
Содержание FLUIDC DHT 1480
Страница 25: ...Fluido DHT Dry Heat Therapy Unit 23 DIAGRAMS Base Assembly 120v and 230v ...
Страница 26: ...Fluido DHT Dry Heat Therapy Unit 24 DIAGRAMS Control Panel Assembly 120v and 230v ...
Страница 27: ...Fluido DHT Dry Heat Therapy Unit 25 DIAGRAMS Lid Assembly 120v and 230v ...
Страница 28: ...Fluido DHT Dry Heat Therapy Unit 26 DIAGRAMS Switch Plate Assembly 120v ...
Страница 29: ...Fluido DHT Dry Heat Therapy Unit 27 DIAGRAMS Switch Plate Assembly 230V ...
Страница 30: ...Fluido DHT Dry Heat Therapy Unit 28 DIAGRAMS Wiring Diagram 120v ...
Страница 31: ...Fluido DHT Dry Heat Therapy Unit 29 DIAGRAMS Wiring Diagram 230v ...
Страница 32: ...Fluido DHT Dry Heat Therapy Unit 30 DIAGRAMS Heater Enclosure Assembly 120v ...
Страница 33: ...Fluido DHT Dry Heat Therapy Unit 31 DIAGRAMS Heater Enclosure Assembly 230v ...
Страница 34: ...Fluido DHT Dry Heat Therapy Unit 32 DIAGRAMS Motor Support Assembly 120v ...
Страница 35: ...Fluido DHT Dry Heat Therapy Unit 33 DIAGRAMS Motor Support Assembly 230v ...
Страница 36: ...Fluido DHT Dry Heat Therapy Unit 34 DIAGRAMS Motor Housing Assembly 120v ...
Страница 37: ...Fluido DHT Dry Heat Therapy Unit 35 DIAGRAMS Motor Housing Assembly 230v ...
Страница 38: ...Fluido DHT Dry Heat Therapy Unit 36 DIAGRAMS Motor Housing Assembly 120v ...
Страница 39: ...Fluido DHT Dry Heat Therapy Unit 37 DIAGRAMS Motor Housing Assembly 230v ...
Страница 42: ...Fluido DHT Dry Heat Therapy Unit 40 DIAGRAMS Final Assembly 120v ...
Страница 43: ...Fluido DHT Dry Heat Therapy Unit 41 DIAGRAMS Final Assembly 230v ...