Fluido DHT
™
Dry Heat Therapy Unit
9
TROUBLESHOOTING
Fluido DHT Model 1480 and 1481
Software Error
Messages
A.
The information provided in this section is
intended to aid in troubleshooting the
Fluido DHT
Model 1480 and 1481 units.
There are only two
software error messages that could indicate a
“Bad Board”. Therefore, should tests contained
in this section indicate the pc board is bad the
board must be replaced as an assembly. No
component level troubleshooting information is
or will be provided by Chattanooga Group for
field troubleshooting of pc board components.
B.
Once the PCB has been determined as bad,
replace the suspected board as described in the
Removal & Replacement section of this manual.
Motor/Blower Error
A.
Should the software detect a problem within the
operation of the Blower/Motor, the display will
show four dashes beneath “AIRSPEED” on the
control panel LCD.
B.
If this indicator is visible, there could be several
possible reasons:
•
The intake filters are clogged and require
cleaning.
•
The motor has overheated and requires a
cooling period before operation is resumed.
•
The Distributor is clogged and requires
changing.
•
The Cellex Medium has entered the heat
cavity of the unit and is restricting the air flow.
Should this indicator be visible, remove the
unit from service immediately. Perform any
and all maintenance and repair necessary
before the unit is placed back into service.
Refer to the appropriate sections of this
manual for proper removal and replacement
procedures.
Temperature Sensor Error
A.
Should the Temperature Sensor experience
problems the software will display four dashes
beneath “Temp” on the LCD.
Should this occur the system may “Lock Up,”
preventing any further operation of the unit. This
error could be caused by several different
problems including:
•
The intake filters are clogged and require
cleaning.
•
The Temperature Sensor is faulty and requires
replacement.
•
The Distributor is clogged and requires
changing.
•
The Cellex Medium has entered the heat
cavity of the unit and is restricting the air flow.
B.
Reset the Temperature Sensor by turning the
unit Off. Press and hold the “Stop” button for
ten to fifteen seconds. Release the “Stop”
button and turn the unit On. It will be necessary
to reset the Clock (refer to Users Manual). If the
Error persists, remove and replace the
Temperature Sensor. If the problem continues to
persist, replace the pc board in the control
panel. Refer to the appropriate section of this
manual for proper removal and replacement
procedures.
B
BLLO
OW
WEER
R//M
MO
OTTO
OR
R PPR
RO
OB
BLLEEM
M
IIN
ND
DIIC
CA
ATTO
OR
R..
TTEEM
MPPEER
RA
ATTU
UR
REE S
SEEN
NS
SO
OR
R
PPR
RO
OB
BLLEEM
M IIN
ND
DIIC
CA
ATTO
OR
R..
Содержание FLUIDC DHT 1480
Страница 25: ...Fluido DHT Dry Heat Therapy Unit 23 DIAGRAMS Base Assembly 120v and 230v ...
Страница 26: ...Fluido DHT Dry Heat Therapy Unit 24 DIAGRAMS Control Panel Assembly 120v and 230v ...
Страница 27: ...Fluido DHT Dry Heat Therapy Unit 25 DIAGRAMS Lid Assembly 120v and 230v ...
Страница 28: ...Fluido DHT Dry Heat Therapy Unit 26 DIAGRAMS Switch Plate Assembly 120v ...
Страница 29: ...Fluido DHT Dry Heat Therapy Unit 27 DIAGRAMS Switch Plate Assembly 230V ...
Страница 30: ...Fluido DHT Dry Heat Therapy Unit 28 DIAGRAMS Wiring Diagram 120v ...
Страница 31: ...Fluido DHT Dry Heat Therapy Unit 29 DIAGRAMS Wiring Diagram 230v ...
Страница 32: ...Fluido DHT Dry Heat Therapy Unit 30 DIAGRAMS Heater Enclosure Assembly 120v ...
Страница 33: ...Fluido DHT Dry Heat Therapy Unit 31 DIAGRAMS Heater Enclosure Assembly 230v ...
Страница 34: ...Fluido DHT Dry Heat Therapy Unit 32 DIAGRAMS Motor Support Assembly 120v ...
Страница 35: ...Fluido DHT Dry Heat Therapy Unit 33 DIAGRAMS Motor Support Assembly 230v ...
Страница 36: ...Fluido DHT Dry Heat Therapy Unit 34 DIAGRAMS Motor Housing Assembly 120v ...
Страница 37: ...Fluido DHT Dry Heat Therapy Unit 35 DIAGRAMS Motor Housing Assembly 230v ...
Страница 38: ...Fluido DHT Dry Heat Therapy Unit 36 DIAGRAMS Motor Housing Assembly 120v ...
Страница 39: ...Fluido DHT Dry Heat Therapy Unit 37 DIAGRAMS Motor Housing Assembly 230v ...
Страница 42: ...Fluido DHT Dry Heat Therapy Unit 40 DIAGRAMS Final Assembly 120v ...
Страница 43: ...Fluido DHT Dry Heat Therapy Unit 41 DIAGRAMS Final Assembly 230v ...