Dry Heat Therapy Unit
3
SAFETY PRECAUTIONS
Precautionary Symbol Definitions
The precautionary instructions found throughout this
manual are indicated by specific symbols.
Understand these symbols and their definitions before
operating or servicing this equipment. The definitions of
these symbols are as follows:
CAUTION
Text with a “CAUTION” indicator will explain possible
safety infractions that could have the potential to
cause minor to moderate injury or damage to
equipment.
WARNING
Text with a “WARNING” indicator will explain possible
safety infractions that will potentially cause serious
injury and equipment damage.
DANGER
Text with a “DANGER” indicator will explain possible
safety infractions that are imminently hazardous
situations that would result in death or serious injury.
EXPLOSION HAZARD
Do not use this equipment in the presence of
flammable anesthetics. This symbol is also
prominently displayed on the serial number plate of
the unit.
NOTE:
Throughout this manual “NOTE” may be found.
The Notes are helpful information to aid in the
particular area or function being described.
Safety Precautions
Read, understand and follow all safety precautions
found in this manual. The following are general safety
precautions that must be read and understood before
attempting any service techniques on these units.
Throughout this manual, specific safety precautions will
be found. Read, understand and follow all safety
precautions.
!
CAUTION
WARNING
!
!
DANGER
•
Read, understand and practice the precautionary and operating
instructions found in this manual. Know the limitations and
hazards associated with using any electrical device. Observe the
precautionary and operational decals placed on the unit.
•
DO NOT operate the unit when connected to any unit other than
Chattanooga Group devices.
•
Refill unit daily to proper fill level with Chattanooga Group Cellex
®
Dry
Heat Medium.
•
Change Cellex Dry Heat Medium every six (6) months.
•
Use only Cellex Dry Heat Medium in the Fluidotherapy units.
•
Clean Inlet Filter(s) daily before unit startup.
•
Use only fingers to operate button controls on the control panel(s).
Use of sharp objects such as pencils or pens will result in damage to
the unit.
•
Turn unit to the “Standby” mode before positioning a patient or
removing a patient from the unit.
•
After or between treatments, do not immediately unplug or turn the
power off from the unit. Allow the unit to process through the “Cool
Down” cycle. The unit cycles into standby mode after treatment time
has elapsed. Standby can be disabled by pressing the
preheat/standby button. The unit goes to a 3 minute cool down mode
after standby is turned off. Turning the power off before cool down
completes is potentially hazardous to the equipment, and could lead
to failure of the unit. It is recommended that power be supplied to the
unit at all times. Keep in mind that the recommended treatment air
speed is 50%
•
Secure all entry ports before turning the unit ON.
•
Check unit temperature before treating patient to ensure correct
temperature.
•
Place the patient in a comfortable position allowing for correct
placement of the limb being treated.
•
Proper storage and transport temperatures for the Fluido DHT
units are 40°F - 158°F (4.5 °C - 70°C). Relative Humidity 85%.
•
This equipment generates, uses and can radiate radio frequency
energy and, if not installed and used in accordance with the
instructions, may cause harmful interference to other devices in
the vicinity. However, there is no guarantee that interference will
not occur in a particular installation. Harmful interference to other
devices can be determined by turning this equipment on and off.
Try to correct the interference using one or more of the following:
Reorient or relocate the receiving device, increase the separation
between the equipment, connect the equipment to an outlet on a
different circuit from that to which the other device(s) are connected
and/or consult the factory field service technician for help.
!
CAUTION
Содержание FLUIDC DHT 1480
Страница 25: ...Fluido DHT Dry Heat Therapy Unit 23 DIAGRAMS Base Assembly 120v and 230v ...
Страница 26: ...Fluido DHT Dry Heat Therapy Unit 24 DIAGRAMS Control Panel Assembly 120v and 230v ...
Страница 27: ...Fluido DHT Dry Heat Therapy Unit 25 DIAGRAMS Lid Assembly 120v and 230v ...
Страница 28: ...Fluido DHT Dry Heat Therapy Unit 26 DIAGRAMS Switch Plate Assembly 120v ...
Страница 29: ...Fluido DHT Dry Heat Therapy Unit 27 DIAGRAMS Switch Plate Assembly 230V ...
Страница 30: ...Fluido DHT Dry Heat Therapy Unit 28 DIAGRAMS Wiring Diagram 120v ...
Страница 31: ...Fluido DHT Dry Heat Therapy Unit 29 DIAGRAMS Wiring Diagram 230v ...
Страница 32: ...Fluido DHT Dry Heat Therapy Unit 30 DIAGRAMS Heater Enclosure Assembly 120v ...
Страница 33: ...Fluido DHT Dry Heat Therapy Unit 31 DIAGRAMS Heater Enclosure Assembly 230v ...
Страница 34: ...Fluido DHT Dry Heat Therapy Unit 32 DIAGRAMS Motor Support Assembly 120v ...
Страница 35: ...Fluido DHT Dry Heat Therapy Unit 33 DIAGRAMS Motor Support Assembly 230v ...
Страница 36: ...Fluido DHT Dry Heat Therapy Unit 34 DIAGRAMS Motor Housing Assembly 120v ...
Страница 37: ...Fluido DHT Dry Heat Therapy Unit 35 DIAGRAMS Motor Housing Assembly 230v ...
Страница 38: ...Fluido DHT Dry Heat Therapy Unit 36 DIAGRAMS Motor Housing Assembly 120v ...
Страница 39: ...Fluido DHT Dry Heat Therapy Unit 37 DIAGRAMS Motor Housing Assembly 230v ...
Страница 42: ...Fluido DHT Dry Heat Therapy Unit 40 DIAGRAMS Final Assembly 120v ...
Страница 43: ...Fluido DHT Dry Heat Therapy Unit 41 DIAGRAMS Final Assembly 230v ...